Chelmsford, MA, United States
ZOLL Medical Corporation recalls
6 recalls recorded since 2018, most recently on 13 February 2025. 4 are listed as ongoing. Status and classification are as published by the FDA.
Recalls6
Ongoing (FDA)4
Terminated or completed2
Class I (FDA)1
Ongoing Class II Medical devices
- Recall started
- 13 February 2025
- Reason
- The potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
- Distribution
- Domestic: Nationwide Distribution International: AE, AI, AM, AR, AT, AU, BD, BE, BG, BH, BM, BN, BS, CA, CH, CL, CN, CO, CR, CZ, DE, DK, DO, EG, ES, FI, FR, GB, GH, GU, HK, HR, HU, ID, IE, IL, IN, IS, IT, JE, JM, JP, KR, KW, KZ, LB, LT, MA, MC, MD, ME, MQ, MS, MT, MX, MY, NC, NG, NL, NO, NZ, OM, PA
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1321-2025, Z-1319-2025, Z-1320-2025, Z-1322-2025, Z-1316-2025, Z-1317-2025, Z-1318-2025, Z-1323-2025
- Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80C; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threate
- Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-01C, (2) G5A-02C, (3) G5A-05C, (4) G5A-08C, (5) G5A-10C, (6) G5A-13C, (7) G5A-15C, (8) G5A-22C, (9) G5A-41C, (10) G5A-85C, (11) G5A-90C (12
- Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-01C, (2) G5S-02C, (3) G5S-10C, (4) G5S-11C; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator
- Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-00A-TSO, (2) G5S-01A, (3) G5S-02A, (4) G5S-05A, (5) G5S-08A, (6) G5S-10A, (7) G5S-11A, (8) G5S-80A-TSO, (9) G5S-80-L (10) G5S-90A; Softwar
- Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-00A, (2) G5A-01A, (3) G5A-02A, (4) G5A-03A, (5) G5A-19A, (6) G5A-80A, (7) G5A-80A-TSO. Software Version: NA Product Description: The Powe
- Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80A, (2) G5A-80A-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for
- Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80C, (2) G5A-80C-TSO. Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for
- Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80A; Software Version: NA Product Description: The Powerheart G5 automated external defibrillator (AED) is designed for treating life-threate
Ongoing Class II Medical devices
- Recall started
- 31 May 2024
- Reason
- G5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest
- Distribution
- Nationwide including PR. Foreign: AU BM CA CH DE DK ES FR GB HK HU ID IE IL IT KR MA MT NL NO PR SA SE SI SN TR
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2324-2024
- ZOLL Powerheart G5 AED, Semi-Automatic, G5Sxxx Family -Automated external defibrillator (AED) is designed for treating life-threatening heart beat irregularities, such as ventricular fibrillation, that cause Sudden Cardi
Ongoing Class I Medical devices
- Recall started
- 30 April 2024
- Reason
- Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy
- Distribution
- Worldwide distribution - US Nationwide and the countries of United Arab Emirates, Argentina, Australia, Azerbaijan, Bangladesh, Belgium, Bahrain, Bolivia, Brazil, Canada, Switzerland, China, Colombia, Czech Republic, Germany, Spain, Great Britain (UK), Greece, Guatemala, Guam, Hong Kong, Croatia,
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2130-2024
- ZOLL 731 Ventilator (EMV+, AEV, Eagle II) - for MRI Compatible Devices - Operator's Guide and Quick Reference Guide (QRGs) as follows: Part Numbers: 1. 799-EGL2-02 2. 799-EGL2-02-01 3. 799-EGL2-02-04 4. 799-EGL2-
Ongoing Class II Medical devices
- Recall started
- 12 June 2019
- Reason
- There is a potential for the device to fail to deliver energy to the patient.
- Distribution
- Worldwide Distribution - US nationwide. The products were distributed to the following foreign countries: Austria, Australia, Bangladesh, Brazil, Canada, China, Denmark, France, Germany, Great Britain, Indonesia, Italy, Japan, Luxemburg, New Zealand, Panama, Russia, Saudi Arabia, Singapore, Spai
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1931-2019
- ZOLL AED PRO Automated External Defibrillator Product Usage: The AED PRO unit is intended to defibrillate victims of ventricular fibrillation or pulseless ventricular tachycardia, for ECG monitoring, and for CPR moni
Terminated Class II Medical devices
- Recall started
- 4 August 2017
- Terminated
- 31 August 2018
- Reason
- A portion of one lot of ZOLL OneStep Complete Electrodes for adults will cause the defibrillator to deliver pediatric energy levels instead of adult energy levels when used with the ZOLL R Series Defibrillator. Defective electrodes will display PEDIATRIC PADS IN USE on the R Series defibrillator. Using the affected product with the M Series Defibrillator poses no risk.
- Distribution
- Worldwide Distribution - US Nationwide in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and in the countries of Canada.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0977-2018
- OneStep CPR Complete, Adult Multi Function Electrodes, Part Numbers: 8900 0214 01 (case of 8 single electrodes, UDI: 10847946016279) and 8900 0224 01 (single electrode, UDI: 00847946016272) Product Usage: Intended Us
Terminated Class II Medical devices
- Recall started
- 30 June 2017
- Terminated
- 9 October 2018
- Reason
- A software anomaly in the 731 software version 05/20/00, was identified, which can lead to a user inadvertently changing device settings.
- Distribution
- Outside US
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0812-2018
- 731 Series Ventilators running software version 05.20.00 The devices in the ZOLL ventilator are indicated for use in the management of infant through adult patients weighing ;:: 5 kg with acute or chronic respiratory
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.