Recall Record

Oranmore, N/A, Ireland

ZIMMER ORTHOPEDICS MANFACTURING LIMITED-GALWAY recalls

1 recall recorded since 2018, most recently on 3 March 2025. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
3 March 2025
Reason
Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.
Distribution
US distribution nationwide. International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.
Initiated
Voluntary: Firm initiated
Recall number
Z-1403-2025

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.