Recall Record

Winterthur, N/A, Switzerland

Zimmer GmbH recalls

27 recalls recorded since 2018, most recently on 20 May 2026. 20 are listed as ongoing. Status and classification are as published by the FDA.

Recalls27
Ongoing (FDA)20
Terminated or completed7
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
20 May 2026
Reason
Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2528-2026, Z-2531-2026, Z-2530-2026, Z-2527-2026, Z-2529-2026

Ongoing Class II Medical devices

Recall started
11 May 2026
Reason
Firm received a complaint with no clinical impact in which a posterior augment was found within the distal augment package.
Distribution
Refer to the attached ZFA 2026-00088_Initial Consignee List and ZFA 2026-00088_Initial Distribution History containing the distribution and the identification of the affected consignees.
Initiated
Voluntary: Firm initiated
Recall number
Z-2578-2026

Ongoing Class II Medical devices

Recall started
30 March 2026
Reason
Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Compromised sterile barrier may lead to clinically insignificant extension of surgery or infection leading to surgical intervention.
Distribution
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MO, MS, NC, NE, NM, NY, OH, OK, PA, SD, TN, TX, UT, WA, WI and the countries of Canada, Australia, New Zealand, Malaysia, EMEA, Chile.
Initiated
Voluntary: Firm initiated
Recall number
Z-1946-2026

Ongoing Class II Medical devices

Recall started
2 December 2025
Reason
Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.
Distribution
US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0972-2026, Z-0970-2026, Z-0975-2026, Z-0968-2026, Z-0982-2026, Z-0974-2026, Z-0971-2026, Z-0969-2026, Z-0980-2026, Z-0976-2026, Z-0983-2026, Z-0978-2026 and 5 more

Ongoing Class II Medical devices

Recall started
26 September 2025
Reason
Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E¿ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
Distribution
US: OH, GA, WI, NE, IA, IL, CA, NY, MD, NC, PA, AL, TN, MI, WA, FL, TX, KS, NM, AZ, and CANADA, AUSTRALIA, NETHERLANDS
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0418-2026, Z-0419-2026

Ongoing Class II Medical devices

Recall started
24 March 2025
Reason
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Distribution
Worldwide distribution - US Nationwide and the countries of Panama, Hong Kong, China.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1648-2025, Z-1646-2025, Z-1645-2025, Z-1649-2025, Z-1647-2025

Ongoing Class II Medical devices

Recall started
27 February 2025
Reason
The "Use with plate 7, 8, 9, 10" compatibility statement on top of the label does not match with the compatibility statements "USE WITH PLATE 5, 6" in the translations section of the same label.
Distribution
Worldwide - US Nationwide distribution in the states pf CA, KS, MA, MD, ME, NE, NJ, OR, SD, and TX. The countries of Canada, China, India, Netherlands, Singapore.
Initiated
Voluntary: Firm initiated
Recall number
Z-1433-2025

Ongoing Class II Medical devices

Recall started
12 February 2025
Reason
There is a potential design issue with the Z1 Offset Stem Inserter leading to fracturing of the distal tip.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1329-2025

Ongoing Class II Medical devices

Recall started
6 January 2025
Reason
It was identified internally that two commingle events occurred where the metal support post within the polyethylene articular surface was assembled incorrectly for three units within the scope.
Distribution
Worldwide - US Nationwide distribution in the states of CA, FL, GA, KS, MO, NE, NJ, PA, TN and the countries of CHINA, INDIA, KOREA, MALAYSIA, NETHERLANDS, SINGAPORE, TAIWAN.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1039-2025, Z-1038-2025

Ongoing Class II Medical devices

Recall started
17 September 2024
Reason
A Gen 5 tail piece was assembled to a Gen 4 Finished Device, resulting in a 0.178" difference in the end of the endoscope slot to the end of the tail piece. This allows the scope to move further into the plastic sheath that runs the length of the device, resulting in an interference fit at the cone, may caus an extension in surgery
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-0285-2025

Ongoing Class II Medical devices

Recall started
8 August 2024
Reason
Mislabeled: Incorrect sleeve in the packaging. The outer packaging is labeled as a standard neck sleeve, however, the product within the box is a -6mm neck sleeve, and vice versa. The neck of the sleeve can be identified by the device etching as either STD or -6
Distribution
Worldwide distribution - US Nationwide and the countries of Japan, Netherlands.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3133-2024, Z-3132-2024

Ongoing Class II Medical devices

Recall started
2 July 2024
Reason
Affected product has an increased risk of postoperative perisprosthetic femoral fracture (PFF). The IFU is being updated to reflect the risk of PFF.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-2450-2024

Ongoing Class II Medical devices

Recall started
28 March 2024
Reason
Out of specification violation of devices that results in a gap existing within the fixture that presses the fiber metal pad to the substrate, causing the pad to not fully bond to the substrate.
Distribution
Worldwide distribution - US Nationwide including the states of AK, AL, CA, FL, GA, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI and the countries of Australia, Canada, Japan, Netherlands.
Initiated
Voluntary: Firm initiated
Recall number
Z-1608-2024

Ongoing Class II Medical devices

Recall started
20 March 2024
Reason
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
Distribution
Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1716-2024, Z-1723-2024, Z-1721-2024, Z-1720-2024, Z-1710-2024, Z-1724-2024, Z-1727-2024, Z-1717-2024, Z-1713-2024, Z-1709-2024, Z-1726-2024, Z-1719-2024 and 9 more

Ongoing Class II Medical devices

Recall started
26 July 2023
Reason
The outer package labeling and product etch are for a Size 6, however, the implant is a Size 4, and vice versa.
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, GA, MI, NJ, PA, TN, and WI. The countries of Brazil, Canada, Netherlands.
Initiated
Voluntary: Firm initiated
Recall number
Z-2519-2023

Ongoing Class II Medical devices

Recall started
27 June 2023
Reason
Thread form issue of the locking holes in that the locking screws would not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock.
Distribution
AR AZ MA ME MO NJ NM NY OR WA Foreign: Brazil China Dominican Republic Korea Netherlamds Taiwan
Initiated
Voluntary: Firm initiated
Recall number
Z-2254-2023

Terminated Class II Medical devices

Recall started
19 May 2023
Terminated
29 May 2026
Reason
Device outer packaging was incorrectly labeled.
Distribution
International distribution in the country of India.
Initiated
Voluntary: Firm initiated
Recall number
Z-2092-2023

Ongoing Class II Medical devices

Recall started
27 February 2023
Reason
Update the compatibility matrix as referred to in the Instructions for Use (IFU) for the CoCr Femoral Head XS- removes the compatibility with the Epsilon Durasul Constrained Acetabular Liners from the matrix due to the range of motion in flexion/extension being less than 100 degrees as recommended per an internationally recognized standard (ISO 21535:2009
Distribution
AL AZ CO FL LA MI OH OK TX Foreign: Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-1294-2023

Ongoing Class II Medical devices

Recall started
7 December 2022
Reason
Two planers were found to be unusable during sawbones shoulder training.
Distribution
US Nationwide distribution in the states of AR, CA, IL, MI, NJ, NY, OK, PA, TX, and WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1098-2023

Ongoing Class II Medical devices

Recall started
6 December 2022
Reason
Zimmer Biomet is conducting a voluntary medical device recall related to the NexGen Stemmed Option Tibial Components due to the clinically and statistically significant higher overall revision rates when these tibial components are used with either the Legacy Posterior Stabilized (LPS) Flex or LPS Flex Gender Solutions Femoral (GSF) components as compared to other total knee arthroplasties in the United Kingdom National Joint Registry (UK NJR). Removing the NexGen Stemmed Option Tibial Component f rom inventory will prevent its future implantation with either the LPS Flex or LPS Flex GSF femoral components and mitigate the increased revision risk with these two specific combinations of tibial and femoral component.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0470-2023, Z-0468-2023, Z-0467-2023, Z-0469-2023, Z-0471-2023, Z-0466-2023

Ongoing Class II Medical devices

Recall started
25 August 2022
Reason
The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant extension of surgery, or infection leading to surgical intervention.
Distribution
Domestic distribution US nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-0007-2023

Terminated Class II Medical devices

Recall started
7 May 2021
Terminated
18 April 2024
Reason
Zimmer Natural Nail (ZNN) System-CMN Femoral Nail potential commingle between CCD angle 130 degree and 125 degree
Distribution
CA, GA, KY, MD Foreign: Australia Brazil, Canada Netherlands
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1792-2021, Z-1793-2021

Terminated Class II Medical devices

Recall started
21 August 2019
Terminated
20 July 2020
Reason
Incorrect labeling on the products package. The label on the package shows size 50-18 whereas the implant inside the packaging size 40-14.
Distribution
Worldwide distribution - iUS Nationwide distribution including states of AL, CO, KS, NJ, NY, OH, and OK, and countries of Belgium, Czech Republic, France, Germany, Switzerland, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2762-2019

Terminated Class II Medical devices

Recall started
8 April 2019
Terminated
11 May 2020
Reason
An investigation identified that the products were possibly manufactured from a different material than defined in the applicable specifications.
Distribution
The products were distributed to the following US states: AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, and WY.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1448-2019, Z-1486-2019, Z-1482-2019, Z-1455-2019, Z-1415-2019, Z-1439-2019, Z-1472-2019, Z-1442-2019, Z-1449-2019, Z-1443-2019, Z-1458-2019, Z-1480-2019 and 64 more

Terminated Class II Medical devices

Recall started
31 October 2018
Terminated
1 April 2020
Reason
Mislabeled Labeled Avenir M¿ller Stem size 1, the stem inside the packaging is an Avenir M¿ller Stem size 2 .
Distribution
International distribution in the countries of Germany and Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-0644-2019

Terminated Class II Medical devices

Recall started
15 March 2018
Terminated
25 April 2019
Reason
Product may contain a gap between the handle and metal chisel which can lead to failure to properly disinfect or sterilize between use.
Distribution
CO, IN, TX, IL
Initiated
Voluntary: Firm initiated
Recall number
Z-1360-2018

Terminated Class II Medical devices

Recall started
28 February 2018
Terminated
25 October 2019
Reason
A potential commingle that could result in the product in the package not matching the product on the outer label.
Distribution
There was no distribution to the US. The distribution occurred to the following countries: Belgium, Switzerland, Germany, France, Canary Islands, Italy, and Thailand.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1525-2018, Z-1524-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.