Recall Record

Westminster, CO, United States

Zimmer Biomet Spine Inc. recalls

2 recalls recorded since 2018, most recently on 16 October 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)1
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
16 October 2023
Reason
Spinal fixation system 7.5mmx 50mm screws are packaged with the correct labeling, but the screws themselves were laser marked with an incorrect size description of 7.5 x 40; use of incorrect screws could lead to screw perforation and indention of vascular or other structures.
Distribution
US Nationwide distribution in the states of MN, NE.
Initiated
Voluntary: Firm initiated
Recall number
Z-0639-2024

Terminated Class II Medical devices

Recall started
30 July 2018
Terminated
28 April 2021
Reason
The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications
Distribution
Distributed Nationwide: AL, CA, CO, FL, GA, HI, IA, ID, IL, IN, LA, MA, MD, MI, MN, NC, NJ, NV, NY, OH, OK, PA, SC, TN, TX, WA and WI.**** International distribution: Australia and EMEA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0868-2019, Z-0866-2019, Z-0867-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.