Recall Record

Thornwood, NY, United States

Zeiss, Carl Inc recalls

4 recalls recorded since 2018, most recently on 24 June 2022. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)0
Terminated or completed4
Class I (FDA)0

Terminated Class III Medical devices

Recall started
24 June 2022
Terminated
7 February 2025
Reason
An incorrect Unique Device Identification (UDI) label was installed on some ZEISS Axio Observer 5 microscopes.
Distribution
US Nationwide distribution in the state of NY.
Initiated
Voluntary: Firm initiated
Recall number
Z-1465-2022

Terminated Class II Medical devices

Recall started
2 October 2019
Terminated
25 July 2024
Reason
When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system
Distribution
Worldwide Distribution - US Nationwide
Initiated
FDA Mandated
Recall number
Z-1629-2020

Terminated Class II Medical devices

Recall started
29 May 2018
Terminated
26 July 2024
Reason
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
Distribution
US Nationwide Distribution
Initiated
FDA Mandated
Recall number
Z-2441-2018

Terminated Class II Medical devices

Recall started
8 May 2018
Terminated
21 April 2020
Reason
In certain eyepiece configurations the laser shutter may not close completely, permitting reflected beams that may be greater than the Class I limit.
Distribution
USA (nationwide)
Initiated
FDA Mandated
Recall number
Z-2058-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.