Xstrahl Limited recalls

3 recalls recorded since 2018, most recently on 27 April 2022. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)1
Terminated or completed2
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
27 April 2022
Reason
There is a potential compatibility issue with the systems and replacement treatment applicators.
Distribution
US Distribution FL, CA, KY
Initiated
Voluntary: Firm initiated
Recall number
Z-1404-2022

Terminated Class II Medical devices

Recall started
27 August 2021
Terminated
4 March 2026
Reason
Base plate may detach from the main body of the treatment applicator.
Distribution
US and ALGERIA, AUSTRALIA, AUSTRIA, BULGARIA, CANADA, COSTA RICA, CROATIA, CYPRUS, DENMARK, FRANCE, GERMANY, HONG KONG, IRELAND, ITALY, KAZAKHSTAN, LATVIA, MALTA, MEXICO, NEW ZEALAND, NORWAY, PANAMA, ROMANIA, RUSSIA, SAUDI ARABIA, SPAIN, SWEDEN, SWITZERLAND, UK, UKRAINE
Initiated
Voluntary: Firm initiated
Recall number
Z-0389-2022

Terminated Class II Medical devices

Recall started
26 April 2021
Terminated
26 January 2024
Reason
If a saved treatment plan with 2 opposing beams is edited prior to approval, then Beam 2 is not updated with the changed parameters upon selecting save, resulting in error messages during the treatment and possible mis-treatment.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, KY, NV, OR, PA, TN, and TX. The countries of Austria, Germany, and Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-0174-2022

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.