Wright Medical Technology, Inc. recalls

22 recalls recorded since 2018, most recently on 18 August 2026. 11 are listed as ongoing. Status and classification are as published by the FDA.

Recalls22
Ongoing (FDA)11
Terminated or completed11
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
18 August 2026
Reason
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the plate's internal threads, which may lead to surgery prolongation to retrieve another device, or alternative surgery method if no replacement device is readily available.
Distribution
US Nationwide distribution in the states of FL, TX, PA, CA, ID, MI, OK, TN, NJ, WI, CO.
Initiated
Voluntary: Firm initiated
Recall number
Z-3334-2026

Ongoing Class II Medical devices

Recall started
6 March 2024
Reason
Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide.
Distribution
US: TX, NE, and AUSTRALIA, BELGIUM, BRAZIL, CANADA, CHINA, COLOMBIA, FRANCE, GERMANY, ITALY, SPAIN, SWEDEN, SWITZERLAND, UK
Initiated
Voluntary: Firm initiated
Recall number
Z-1477-2024

Ongoing Class II Medical devices

Recall started
13 June 2023
Reason
The devices were manufactured with the pin sleeve holes too narrow, which does not allow the pin sleeves to pass through the pin sleeve holes, thereby preventing pin placement through the pin holes.
Distribution
There was no U.S. distribution, including government and military. Foreign distribution was made to Australia, Belgium, Canada, Chile, China, Colombia, France, Germany, Italy, Mexico, Netherlands, Spain, Sweden, Switzerland, Taiwan, and the United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2105-2023

Ongoing Class II Medical devices

Recall started
10 April 2023
Reason
Two lots of Ortholoc 3Di Fusion plates contain incorrect dorsiflexion descriptions laser marked on the parts.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1467-2023, Z-1466-2023

Ongoing Class II Medical devices

Recall started
2 February 2023
Reason
Stryker has identified an issue that impacts two specific lots of CLAW" II plates. The plates within these two lots were found to have been comingled, resulting in the incorrect product in packages. Packaging for 402S0425 incorrectly contained 40241425, and vice versa.
Distribution
US Nationwide distribution in the state of NY.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1170-2023, Z-1169-2023

Terminated Class II Medical devices

Recall started
30 September 2022
Terminated
21 August 2026
Reason
The tibial tray lock detail is oversized (larger than specification).
Distribution
US Nationwide distribution in the state of New York.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0139-2023, Z-0140-2023

Ongoing Class II Medical devices

Recall started
3 August 2022
Reason
The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm +2 stem.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0006-2023, Z-0005-2023

Ongoing Class II Medical devices

Recall started
27 July 2022
Reason
An incorrect sizing label was applied to the carton packaging.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, GA, NY, PA, and TN. The countries of the Australia, France, the Netherlands, Sweden, and the United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1567-2022

Ongoing Class II Medical devices

Recall started
14 June 2022
Reason
The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use
Distribution
Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto Rico
Initiated
Voluntary: Firm initiated
Recall number
Z-1366-2022

Ongoing Class II Medical devices

Recall started
2 June 2022
Reason
The instrument is incorrectly color coded.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, FL, GA, IA, IL, MA, MI, MN, MO, NC, NY, OH, PA, TX, and WA. The countries of Australia, Belgium, Canada, France, Germany, Hong Kong, Israel, Sweden, Switzerland, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-1291-2022

Terminated Class II Medical devices

Recall started
10 March 2022
Terminated
1 April 2026
Reason
The drill and adjustable drill guide can jam/bind intraoperatively resulting in the drill guide breaking.
Distribution
US distribution to Florida, North Carolina, Ohio, and Texas
Initiated
Voluntary: Firm initiated
Recall number
Z-0853-2022

Terminated Class II Medical devices

Recall started
15 February 2022
Terminated
5 October 2022
Reason
An incorrect screw is packaged in the product, which contains a locking screw instead of a non-locking screw.
Distribution
International distribution to the country of France.
Initiated
Voluntary: Firm initiated
Recall number
Z-0786-2022

Terminated Class II Medical devices

Recall started
9 February 2022
Terminated
25 January 2024
Reason
Product was labelled with the incorrect manufacturing and distribution dates.
Distribution
International distribution in the countries of Japan and the United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0679-2022

Ongoing Class II Medical devices

Recall started
3 December 2021
Reason
The incorrect product is contained in the packaging.
Distribution
Worldwide distribution in the states of AZ, IA, ID, NC, NY, VA, and WV. The countries of Canada, France, Germany, Lebanon, The Netherlands, Saudi Arabia, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0393-2022, Z-0392-2022

Ongoing Class II Medical devices

Recall started
3 December 2021
Reason
Incorrect product is contained in the packaging.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, IA, and SC. The countries of Australia, Canada, Chile, Spain, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0401-2022

Terminated Class II Medical devices

Recall started
4 November 2021
Terminated
6 January 2023
Reason
The package contents and package labeling do not match.
Distribution
US Nationwide distribution in the states of MI and PA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0316-2022, Z-0315-2022

Terminated Class II Medical devices

Recall started
26 August 2020
Terminated
27 April 2021
Reason
Lack of sterility assurance
Distribution
USA states: CA, CO, FL, GA, MA, MO, OH, PA, TX OUS: Australia, Canada, Chile, Israel, Spain
Initiated
Voluntary: Firm initiated
Recall number
Z-2989-2020

Terminated Class II Medical devices

Recall started
28 July 2020
Terminated
16 November 2022
Reason
Hardware removal kit does not include drivers to remove Dart-Fire 2.5 and 3.0mm Headless screws.
Distribution
US Nationwide distribution including in the states of CO, TX, OR, KX, SC, CA, AR, NJ, UT, VA, MO, NC, FL, IL.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2797-2020, Z-2796-2020

Terminated Class II Medical devices

Recall started
9 April 2020
Terminated
2 November 2020
Reason
One lot of INBONE Tibial Trays is missing the plasma coating.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1924-2020

Terminated Class II Medical devices

Recall started
27 December 2019
Terminated
23 July 2021
Reason
Two lots of EVOLVE Modular Radial Head could have the wrong size product in the package. Both the label and laser marking are incorrect on the affected product.
Distribution
US
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1101-2020, Z-1102-2020

Terminated Class II Medical devices

Recall started
14 March 2018
Terminated
5 October 2020
Reason
Tow lot of VALOR(TM) Nail could potentially be missing the assembled compression screw.
Distribution
US and Argentina, Belgium, Brazil, Chile, Colombia, Germany, Italy, Mexico, Peru, Thailand, and United Kingdom
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1375-2018, Z-1376-2018

Terminated Class II Medical devices

Recall started
8 November 2016
Terminated
3 November 2020
Reason
The firm discovered a lack of adequate documentation confirming sterility on certain products.
Distribution
US and Norway, Italy, Germany, France, South Africa, UK, Taiwan, Austria, Japan, Turkey, Denmark, Canada, Switzerland, Belgium, The Netherlands, Spain
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0564-2020, Z-0568-2020, Z-0570-2020, Z-0565-2020, Z-0573-2020, Z-0569-2020, Z-0572-2020, Z-0562-2020, Z-0566-2020, Z-0567-2020, Z-0563-2020, Z-0571-2020