Westmed, Inc. recalls

3 recalls recorded since 2018, most recently on 27 September 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)1
Terminated or completed2
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
27 September 2024
Reason
Due to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may leak or disconnect and may cause life threating conditions or death.
Distribution
U.S.: AK, CA, CO, DC, FL, GA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OH, OR, PA, SC, TX, VA, VT, WA, and WV O.U.S.: N/A
Initiated
Voluntary: Firm initiated
Recall number
Z-0263-2025

Terminated Class II Medical devices

Recall started
11 November 2020
Terminated
7 February 2023
Reason
There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0746-2021, Z-0745-2021, Z-0743-2021, Z-0741-2021, Z-0744-2021, Z-0742-2021

Terminated Class II Medical devices

Recall started
15 January 2019
Terminated
5 February 2021
Reason
A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of CA, GA, IL,OH, MA, MI, MO, NC, OH, OR, TX, VA, and WI. The country of Spain.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2435-2020, Z-2434-2020, Z-2436-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.