WELCH ALLYN, INC/MORTARA recalls
2 recalls recorded since 2018, most recently on 11 November 2020. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 11 November 2020
- Terminated
- 8 August 2023
- Reason
- The radio within the device can become disassociated with the wireless access point.
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IN, KS, KY, MA, MD, MI, MO, MT, NC, ND, NH, NM, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, WA, WI, WV, WY and the countries of Australia, Canada, Czech, Finland, GB, Germany, Italy, Kuwai
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0553-2021
- ELI 380 Electrocardiograph - Product Usage: intended to be a high-performance, multichannel resting electrocardiograph. As a resting electrocardiograph, the ELI 380 simultaneously acquires data from each lead.
Terminated Class II Medical devices
- Recall started
- 18 March 2020
- Terminated
- 9 February 2026
- Reason
- Impacted Welch Allyn products may not meet the Defibrillation Withstand requirements of EN/IEC 60601-2-25 Medical Electrical Equipment.
- Distribution
- US nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, ON, OR, PA, PR, RI, SC, SD, TN,UT, VA, VT, WA, WI, WV and WY.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1814-2020, Z-1810-2020, Z-1816-2020, Z-1819-2020, Z-1813-2020, Z-1818-2020, Z-1817-2020, Z-1811-2020, Z-1815-2020, Z-1812-2020
- X12+ and patient cable for the X12+ Telemeter with the following device name: PAT CBL 10WIRE LEADFORM AHA SNAP, PAT CBL 10WIRE LEADFORM IEC SNAP, PAT CBL 10WIRE LEADFORM XL AHA SNAP, PAT CBL 10WIRE LEADFORM XL IEC SNAP,
- Lead diagnostic electrocardiograph under the following device names: ELI 10, ELI 350, ELI 150c, ELI 250c, ELI 230, ELI 280, ELI 380, R-Scribe and Connex Cardio. Product Usage: A non-invasive prescription device indi
- Surveyor S12/S19 and patient cable for the S12/S19 Bedside Monitor under the following device names: PAT CBL 10WIRE AHA SNAP JSCREW, PAT CBL 10WIRE IEC SNAP JSCREW. The Surveyor S12 patient monitor is designed for co
- Replaceable lead set for the Wireless Acquisition Module (WAM), which is the ECG front-end for various Electrocardiographs. Sold under the following device names: LEAD SET WAM 10 WIRE BANANA AHA GRAY, LEAD SET WAM 10
- T12. Incorporates wireless electrocardiographic technology to achieve the real-time acquisition and RF transmission of simultaneous 12-lead ECG data with diagnostic quality while allowing the patient to be ambulatory.
- Q-Tel. The Q-Tel RMS system is a computer-based cardiac and pulmonary rehabilitation data acquisition and editing system.
- Q-Stress/XScribe. Q-Stress or XScribe is a diagnostic device capable of real time ECG display, heart rate measurement, ST analysis and ventricular ectopic beat detection using wired or wireless acquisition modules.
- ELI PC. The ELI PC Service automatically converts resting ECG recordings collected with WAM PC patient cables and WebUpload into DICOM 12-lead ECG Waveform files.
- and 2 more product lines
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.