WASATCH PHOTONICS recalls

1 recall recorded since 2018, most recently on 4 August 2025. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
4 August 2025
Reason
Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.
Distribution
U.S.
Initiated
FDA Mandated
Recall number
Z-2478-2025