Vyaire Medical recalls

12 recalls recorded since 2018, most recently on 10 April 2025. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls12
Ongoing (FDA)3
Terminated or completed9
Class I (FDA)9

Ongoing Class I Medical devices

Recall started
10 April 2025
Reason
Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.
Distribution
Worldwide distribution - US Nationwide and the countries of Germany, Poland, Belgium, Spain, Romania, UK, Italy, Netherlands, India, Slovenia, Qatar, and Saudi Arabia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1697-2025, Z-1698-2025

Ongoing Class I Medical devices

Recall started
8 April 2024
Reason
Potential of the nozzle separating during patient use.
Distribution
US nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1730-2024

Ongoing Class I Medical devices

Recall started
10 January 2024
Reason
Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
Distribution
Worldwide distribution. US nationwide, Brazil, Canada, Switzerland, France, UK, Mexico, Netherlands, Saudi Arabia, Sweden, and South Africa
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1062-2024, Z-1064-2024, Z-1061-2024, Z-1060-2024, Z-1059-2024, Z-1063-2024, Z-1058-2024, Z-1057-2024, Z-1065-2024

Completed Class II Medical devices

Recall started
27 October 2023
Reason
Two patient circuits, 29028-003 and 29028-004, for the 3100 High Frequency Oscillatory ventilator that are only approved for use outside the U.S. were distributed in the U.S.
Distribution
US: Florida
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0489-2024, Z-0488-2024

Completed Class II Medical devices

Recall started
5 May 2023
Reason
Vyaire Medical identified two patient safety risks during the use of the bellavista 1000 and bellavista 1000e: 1. Under certain conditions of use, the touchscreen may become unresponsive and the device application stops responding (APP Hang). When this occurs, if the user continues to interact with the touchscreen, a message will appear that states DeviceSoftware.Application is not responding or a decommission screen will appear on the user interface. The ventilator will issue both an audible and visual alarm. Ventilation continues without interruption with the settings applied prior to the APP Hang error. 2. The potential for a use error has been identified if the operator applies the proposed settings without confirming the settings are suitable for the patient.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1967-2023, Z-1966-2023

Terminated Class I Medical devices

Recall started
20 December 2021
Terminated
28 August 2024
Reason
Potential cessation in ventilation can occur under specific conditions.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL, CA, CO, CT, DC, FL, GA, ID, IL, IN, LA, MA, MD, MI, MO, MS, NC, NJ, NY, OH, OK, PA, SD, TX, VA, WA, WI and WV, including Puerto Rico and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Bosnia and Herzegovina,
Initiated
Voluntary: Firm initiated
Recall number
Z-0622-2022

Terminated Class II Medical devices

Recall started
25 November 2019
Terminated
8 August 2023
Reason
The G6 bellavista 1000 US ventilators may experience intermittent failures: Lack of acoustic high priority alarm, presence of a 'no alarm' condition, or presence of non-responsive touch screen.
Distribution
US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, MD, MI, PA, TX, WA and Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall number
Z-1439-2020

Terminated Class I Medical devices

Recall started
13 March 2019
Terminated
22 September 2020
Reason
Testing has demonstrated aluminum elution from the enFlow Disposable Cartridge during fluid warming.
Distribution
Worldwide Distribution. US nationwide, Brazil, Canada, Chile, Colombia, Ecuador, Hong Kong, Hungary, Israel, India, Jordan, Korea, Kuwait, Kazakhstan, Lebanon, Macedonia, Mexico, Malaysia , Myanmar, Netherlands, New Zealand, Oman, Peru, Qatar, Russian Federation, Saudi Arabia, Thailand, Turkey, Uni
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1385-2019, Z-1386-2019

Terminated Class I Medical devices

Recall started
20 June 2018
Terminated
28 July 2020
Reason
Vyaire Medical has discovered potential patient safety risk with associated with the Tri-Flow Subglottic Suction System. The distal soft tip of the catheter has been identified as having the potential to break off during patient use and result in aspiration of a solid foreign body.
Distribution
Domestic Distribution: AZ, CA, CT, FL, GA, IA, IL, IN, MA, MD, MN, NJ, OH, OK, SD, TX, WA, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2826-2018

Terminated Class I Medical devices

Recall started
8 May 2018
Terminated
7 July 2020
Reason
An error in manufacturing of the AirLife Resuscitation Devices that has the potential to obstruct airflow of the supplemental oxygen delivery during use.
Distribution
US distribution only to: AL, AR, AZ, CA, CA, CO, FL, GA, HI, IA, IL, KY, LA, MA, MD, MI, MN, MO, MS, NC,NJ, NY, OH, OR, PR, SD, TX, VA, WA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2552-2018, Z-2553-2018

Terminated Class I Medical devices

Recall started
27 April 2017
Terminated
9 May 2019
Reason
The mask component on various lots of the resuscitation devices have been identified as having the potential to exhibit a sticking condition which may result in a difficult to/unable to disconnect condition of the mask from the elbow of the resuscitator.
Distribution
Worldwide Distribution - US (Nationwide) and in the countries of Canada, Puerto Rico, United Arab Emirates, Philippines, and Europe.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1580-2018, Z-1549-2018, Z-1551-2018, Z-1560-2018, Z-1567-2018, Z-1557-2018, Z-1569-2018, Z-1566-2018, Z-1572-2018, Z-1579-2018, Z-1570-2018, Z-1564-2018 and 23 more

Terminated Class I Medical devices

Recall started
29 November 2014
Terminated
5 February 2021
Reason
Manufacturing error in humidification chamber may cause an overfill/overflow condition of the chamber. The may case a water back-up into the patient circuit resulting in an excessive amount of water entering the airway/lungs of the ventilated patient.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0346-2018, Z-0343-2018, Z-0344-2018, Z-0345-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.