10 recalls recorded since 2018, most recently on 6 August 2026. 6 are listed as ongoing. Status and classification are as published by the FDA.

Recalls10
Ongoing (FDA)6
Terminated or completed4
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
6 August 2026
Reason
Software system may associate image data from one patient with another patient's record, potentially resulting in incorrect diagnosis or treatment.
Distribution
Worldwide - US Nationwide distribution in the states of MA, MN, TX, AL, GA, CA and the countries of Belgium, Denmark, Germany, Netherlands, Poland, Spain, Sweden, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-3276-2026

Ongoing Class II Medical devices

Recall started
13 June 2025
Reason
Reports of IVUS catheter and guide wire entanglement during radial-to-peripheral procedures, due to catheter use without appropriate sheath and/or guide catheter, which may require further intervention, such as surgical removal; Firm is restating instructions for use and adding: "Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire"
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries
Initiated
Voluntary: Firm initiated
Recall number
Z-2221-2025

Terminated Class II Medical devices

Recall started
16 May 2022
Terminated
23 December 2024
Reason
Due to a potential failed sterilization process.
Distribution
U.S.: WA and WI O.U.S.: Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-1270-2022

Ongoing Class III Medical devices

Recall started
13 May 2022
Reason
Digital IVUS catheter instructions for use ( IFU) do not contain contrainidications
Distribution
Worldwide distribution - US Nationwide distribution in the states of NY, CA, AZ, LA, NV, AL, MI, WI, CO, TX, NJ, OR, FL, KS, IL, MD, PA, OH, NC, OK, MN, AK, GA, MO, SC, MS, KY, AR, VA, SD, TN, MA, DE, NE, MT, CT, WV, WA, IN, IA, HI, NH, WY, UT, ID, DC, RI, NM, ME, VT and the countries of ES, SE, F
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1301-2022, Z-1302-2022

Ongoing Class III Medical devices

Recall started
3 February 2022
Reason
Multi-Modality Touch Screen Modules (MM-TSM), used with integrated IVUS systems, may crack, or become damaged, which may affect their usability. They could become non-functional. Therefore, bottom protective beams will be replaced with full protective metal brackets.
Distribution
US: AZ, IL, NV, MI, FL, NC, AK, PA, WI, MN, CA, AR, NJ, TX, OR, NM, SD, CO, KY, LA, DE, MA, MT, MO, VA, AL, NY, WV, SC, IN, GA, IA, OH, UT, NH, MS, TN, OK, ID, MD, HI, RI, WY, DC, WA, KS, NE, ND, CT, VT. OUS: SK, IT, DK, GB, DE, SE, IN, KW, QA, PL, NL, CL, JP, BE, AU, FR, PF, PT, NZ, CA, ES, PK, CY
Initiated
Voluntary: Firm initiated
Recall number
Z-0917-2022

Ongoing Class III Medical devices

Recall started
24 January 2022
Reason
Multi-Modality Touch Screen Modules (MM-TSM), which are bedside/secondary control units, used as part of Mobile IVUS systems, are missing the following labeling information: Unique Device Identifier, Legal Manufacturer Information, Part Number, Rx Symbol, and Model Reference Information.
Distribution
US: CO, GA, MA, TX, AZ, NY, PA, CA, FL, MO, OR, MD, NC, TN OUS: JP, DE, GB, NL, IT
Initiated
Voluntary: Firm initiated
Recall number
Z-0901-2022

Ongoing Class II Medical devices

Recall started
19 January 2022
Reason
If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR Co-Registration results to be displayed, that may cause users to mistakenly use incorrect measurements, leading to inappropriate patient treatment. Image acquisition and processing system operators manual to be updated to include workflow alternatives.
Distribution
US: WI, CA, CO, MO, ID, DC, LA, VA, KS, OK, GA, KY, IL, AZ, NY, IN, FL, TX, NC, OH, WA, OR, NM, MT, IA, MI, TN, MN, PA, AL, MA, WV, NV, NH, CT, NE, NJ, HI, AR, UT, MD, SC, MS, AK, DE OUS: JP, SA, GB, DE, PT, SG, PL, ZA, NL, MT, ES, IN, GR, EG, FI, HK, FR, NO, DK, BE, AT, IT, CO, KW, MX, PS, QA, C
Initiated
Voluntary: Firm initiated
Recall number
Z-0885-2022

Terminated Class II Medical devices

Recall started
7 November 2018
Terminated
14 July 2022
Reason
Interoperability issue that affects certain systems that use the recalled software when it is connected to specific Hemodynamic Systems (HMS).
Distribution
Distribution was nationwide. There was government/military distribution. Foreign distribution was made to Canada, Austria, Australia, Badajoz, Bahrain, Belgium, Brazil, Bulgaria, Chile, China, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong K
Initiated
Voluntary: Firm initiated
Recall number
Z-0674-2020

Terminated Class II Medical devices

Recall started
17 October 2018
Terminated
30 April 2021
Reason
Visions PV .035 Intravascular Ultrasound (IVUS) Catheters (catalog number 88901 adnd81234) due to the potential inability to pass the guidewire through the y-port of the catheter.
Distribution
Worldwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0726-2019, Z-0727-2019

Terminated Class II Medical devices

Recall started
22 March 2018
Terminated
28 April 2021
Reason
The incorrect resistor has been used in affected CORE M2 Imaging Systems, which causes noise and results in an overly bright IVUS Image.
Distribution
Worldwide Distribution -- U.S., to the states of California, Connecticut, Tennessee, North Carolina, Maryland, and Oregon; and, the country of Australia.
Initiated
Voluntary: Firm initiated
Recall number
Z-1891-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.