Lincoln, NE, United States
VIRTUAL INCISION CORPORATION recalls
1 recall recorded since 2018, most recently on 23 April 2025. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Completed Class II Medical devices
- Recall started
- 23 April 2025
- Reason
- The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of the Surgical Minibot arms.
- Distribution
- US Distribution to States: CA, FL, NE
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1905-2025
- MIRA SURGICAL SYSTEM, Table mounted miniaturized electromechanical surgical system, Catalog Number MIRA-CAP-110, Software Version 4.0.3
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.