Villa Sistemi Medicali S.P.A. recalls
1 recall recorded since 2018, most recently on 4 November 2019. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 4 November 2019
- Reason
- An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, and Puerto Rico.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0763-2020, Z-0766-2020, Z-0765-2020, Z-0764-2020
- Juno DRF, Model Number 709020 - Product Usage: General. multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures.
- Apollo EZ, Model Number 9784152036 - Product Usage: The Apollo EZ remote-controlled table is X-ray equipment that can be used to perform general radiographic and fluoroscopic examinations. The system is a fixed position
- Apollo DRF, Model Number 9784220831 - Product Usage: The remote controlled table Apollo DRF is an X-ray equipment that allows to perform general radiographic and fluoroscopic examinations. The system is a fixed position
- Apollo, Model Numbers 9784100231, 9784120131, 9784130131 - Product Usage: Apollo table is a remote controlled X-ray system suitable for radiography and fluoroscopy examinations. The system is stationary mounted. The move