Vero Biotech, LLC recalls

3 recalls recorded since 2018, most recently on 17 November 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)1
Terminated or completed2
Class I (FDA)2

Ongoing Class II Medical devices

Recall started
17 November 2023
Reason
Cassette did not meet an 1n-process flow criterion.
Distribution
US: AZ and FL
Initiated
Voluntary: Firm initiated
Recall number
Z-0564-2024

Completed Class I Medical devices

Recall started
9 September 2021
Reason
Manufacturing defect involving reversal of the inlet and outlet gas lines, resulting in delivery of nitrogen dioxide (NO2) instead of nitric oxide (NO).
Distribution
Distributed within US to Texas and Illinois.
Initiated
Voluntary: Firm initiated
Recall number
Z-2587-2021

Terminated Class I Medical devices

Recall started
19 May 2021
Terminated
12 October 2023
Reason
Reports received of NO measured below desired dose during transition between primary console and backup console since Software Version 2.2.3 was uploaded to consoles in the field.
Distribution
Distributed nationwide to TX, MD, VA, GA, FL, KS, LA, MN, and NC
Initiated
Voluntary: Firm initiated
Recall number
Z-1863-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.