Ventec Life Systems, Inc. recalls

6 recalls recorded since 2018, most recently on 28 July 2026. 6 are listed as ongoing. Status and classification are as published by the FDA.

Recalls6
Ongoing (FDA)6
Terminated or completed0
Class I (FDA)3

Ongoing Class II Medical devices

Recall started
28 July 2026
Reason
Labeling; Oxygen Cylinders do not contain required medical gas labeling
Distribution
US Nationwide distribution in the states of FL, IA, OH, NE, TN.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3223-2026, Z-3222-2026, Z-3221-2026

Ongoing Class II Medical devices

Recall started
8 May 2026
Reason
Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
Distribution
US Nationwide distribution in the state of UT.
Initiated
Voluntary: Firm initiated
Recall number
Z-2525-2026

Ongoing Class I Medical devices

Recall started
25 March 2026
Reason
Respiratory system intended to provide continuous/intermittent ventilator support, may not have been evaluated under all intended high-pressure conditions during production testing, so an oxygen leak condition could occur that may lead to reduced below intended levels of fraction of inspired oxygen delivery prior to and during ventilation and this oxygen-enriched environment may increase fire risk
Distribution
US: OH, GA, CA, MO
Initiated
Voluntary: Firm initiated
Recall number
Z-1942-2026

Ongoing Class I Medical devices

Recall started
3 February 2025
Reason
Multi-Function Ventilators were serviced using incorrect parts which have the potential to cause unexpected shutdown, or when using an active circuit could result in inaccurate tidal volume monitor, not triggering on patient efforts, less inspiratory volume, less inspiratory pressure, and less PEEP delivered.
Distribution
US distribution in CA, NY, and MO.
Initiated
Voluntary: Firm initiated
Recall number
Z-1196-2025

Ongoing Class I Medical devices

Recall started
9 January 2024
Reason
The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
Distribution
U.S.: CA, CO, FL, HI, IN, MO, NC, and WA O.U.S.: Japan
Initiated
Voluntary: Firm initiated
Recall number
Z-1070-2024

Ongoing Class II Medical devices

Recall started
10 May 2021
Reason
Due to a component failure, ventilators may unexpectedly shut down or have a loss of blower function while in patient use. The shutdown potentially prevents the ventilator from providing ventilation therapy.
Distribution
Distribution US nationwide, Hong Kong, and Japan
Initiated
Voluntary: Firm initiated
Recall number
Z-1902-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.