United Orthopedic Corp. recalls

4 recalls recorded since 2018, most recently on 20 May 2026. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)3
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
20 May 2026
Reason
Due to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
Distribution
US Nationwide distribution in the states of Texas and Michigan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2885-2026, Z-2883-2026, Z-2884-2026, Z-2886-2026

Ongoing Class II Medical devices

Recall started
19 March 2026
Reason
Due to a design issue causing the incorrect engraved numbering on implant that could result in the incorrect orientation of the implant.
Distribution
US Nationwide distribution in the states of Texas and Michigan.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2897-2026, Z-2898-2026

Ongoing Class II Medical devices

Recall started
3 January 2024
Reason
Their is a potential that the implant curved stem may be oriented incorrectly.
Distribution
US: CA,KY, MI, IN, TN, CO, FL, TX, NV, OUS: Switzerland, United Kingdom, Malaysia, Philippines, Taiwan
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1217-2024, Z-1215-2024, Z-1216-2024

Terminated Class II Medical devices

Recall started
24 November 2017
Terminated
9 October 2018
Reason
UOC received a customer compliant report from our distributor on October 31, 2017 which stated that contents of the package do not match the product labeling.
Distribution
Worldwide Distribution - US Distribution and to the countries of: UK, Iran, Greece, Pakistan, Switzerland, France, China, Taiwan, Italy, Korea and South Africa.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1128-2018, Z-1129-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.