Recall Record

Ratisbon, N/A, Germany

UFSK - International OSYS Gmbh recalls

1 recall recorded since 2018, most recently on 31 May 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
31 May 2023
Reason
Service life of the electrical seat part drive of the treatment chair may be shortened unexpectedly after several years of use due to a potential wear-related reduction in the self-locking of the drive and may remain undetected. A worsened self-locking of the seat drive can lead to overriding the drive s end position when positioning the patient with the seat part in an inclined position
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2212-2023

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.