Trumpf Medical Systems, Inc. recalls

1 recall recorded since 2018, most recently on 15 December 2017. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
15 December 2017
Terminated
23 April 2020
Reason
The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.
Distribution
Worldwide Distribution - US Distribution to the states of : GA, IL, IN, PA, SC, TX, UT, VA. and to the countries of : Austria, Australia, Bangladesh, Belgium, France, Switzerland, Czech Republic, Germany, Denmark, Spain, United Arab Emirates, United Kingdom, Ireland, Italy, Kuwait, Lithuania, Pol
Initiated
Voluntary: Firm initiated
Recall number
Z-0824-2018