Charleston, SC, United States
Trumpf Medical Systems, Inc. recalls
1 recall recorded since 2018, most recently on 15 December 2017. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 15 December 2017
- Terminated
- 23 April 2020
- Reason
- The radiated emission from the TruVidia" Wireless Receiver is above the normative accepted limits, in which the radiated emission may interact with other devices in the operating room and cause them not to function properly.
- Distribution
- Worldwide Distribution - US Distribution to the states of : GA, IL, IN, PA, SC, TX, UT, VA. and to the countries of : Austria, Australia, Bangladesh, Belgium, France, Switzerland, Czech Republic, Germany, Denmark, Spain, United Arab Emirates, United Kingdom, Ireland, Italy, Kuwait, Lithuania, Pol
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0824-2018
- Camera System, Television, Surgical without audio Intended for transmission and documentation of surgical interventions for the purpose of archiving and teaching.
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.