10 recalls recorded since 2018, most recently on 19 March 2024. 4 are listed as ongoing. Status and classification are as published by the FDA.

Recalls10
Ongoing (FDA)4
Terminated or completed6
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
19 March 2024
Reason
Nail system contains an out of specification driver-socket which could result in inability to pick up nails or apply sufficient torque.
Distribution
US Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IL, LA, MA, MI, NC, NJ, NV, OH, OK, PA, TN, TX, UT, VA, WA, WI.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1792-2024, Z-1791-2024

Ongoing Class II Medical devices

Recall started
28 November 2023
Reason
Plates, part of a wrist fixation system, are affixed during surgery with bearings that may pop out from distal peg holes.
Distribution
US: FL, MI, AL, NY, CA
Initiated
Voluntary: Firm initiated
Recall number
Z-0932-2024

Ongoing Class II Medical devices

Recall started
22 December 2022
Reason
Overly aggressive tension on the pull suture, while passing the medial button through a bone/soft tissue tunnel across the fibula and tibia, can exceed its strength causing it to break.
Distribution
US Nationwide distribution in the states of IL, PA, TX, GA, TN, AL, MO, FL, AR, LA, HI, OH, TX, KS
Initiated
Voluntary: Firm initiated
Recall number
Z-1101-2023

Ongoing Class II Medical devices

Recall started
28 September 2018
Reason
The epoxy on Apex manufactured epoxy-coated instruments can pop off.
Distribution
Distribution was nationwide. There was military distribution but no government distribution. Foreign distribution was made to Australia, China, South Africa, and Sweden.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0355-2020, Z-0358-2020, Z-0356-2020, Z-0357-2020, Z-0359-2020

Terminated Class II Medical devices

Recall started
24 August 2018
Terminated
28 December 2020
Reason
The bone screw is shorter than the intended design specification.
Distribution
US Distribution to states of FL and MO.
Initiated
Voluntary: Firm initiated
Recall number
Z-1214-2019

Terminated Class II Medical devices

Recall started
24 August 2018
Terminated
28 April 2021
Reason
Modifications to the specification were not appropriately qualified through the company's design control process which may result in structural integrity issues of the device.
Distribution
Distribution was made to CA, FL, MD, TX, and WA. There was military and foreign distribution was made to Australia. There was no government distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0118-2020

Terminated Class II Medical devices

Recall started
15 August 2018
Reason
The firm has become aware that the angled-hole of certain semi-tubular design bone plates may have been modified from its specification, and the modification may not have been fully qualified.
Distribution
US Nationwide distribution in the states of MD, UT.
Initiated
Voluntary: Firm initiated
Recall number
Z-0366-2020

Terminated Class II Medical devices

Recall started
16 April 2018
Terminated
28 January 2021
Reason
Two instrument holes located on the side of the plate, which act to align the saw guide are located on the wrong side of the plate
Distribution
Worldwide Distribution in the state of TX and countries of Australia, China Sweden
Initiated
Voluntary: Firm initiated
Recall number
Z-2877-2018

Terminated Class II Medical devices

Recall started
13 April 2018
Terminated
22 January 2021
Reason
During a retrospective review of sterilization records, TriMed has determined that evidence of quarterly dose audits/verification required to comply with the regulations could not be found. Sterilization records indicate that all lots of staples were irradiated to at least the defined dose. There is a risk that this issue could result in patient infection if not properly sterilized.
Distribution
U.S.: IL, UT, PA, NJ, WA, TX, NV, CA, NC, MS, KS, MO, MN, FL; OUS: Australia and Sweden
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2264-2018, Z-2263-2018, Z-2262-2018, Z-2265-2018

Terminated Class II Medical devices

Recall started
11 January 2018
Terminated
18 January 2024
Reason
Following the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate down the shaft and/or potentially onto the sterilization tray or caddy.
Distribution
US, China, Latin America
Initiated
Voluntary: Firm initiated
Recall number
Z-1373-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.