Recall Record

Tustin, CA, United States

Toshiba American Medical Systems Inc recalls

2 recalls recorded since 2018, most recently on 4 January 2018. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls2
Ongoing (FDA)0
Terminated or completed2
Class I (FDA)0

Terminated Class II Medical devices

Recall started
4 January 2018
Terminated
10 October 2023
Reason
The power supply unit may be damaged and may fail to start up when the power plug is inserted into a power outlet while the main power switch on the rear of the system is set to ON.
Distribution
USA (nationwide) Distribution to the states of : CA, CO, DE, GA, HI, IA, ID, IL, IN, MD, MI, MN, MO, NC, ND, NV, OH, OR, PA, SC, SD, TN, TX, VA, WA, WV and WY.
Initiated
Voluntary: Firm initiated
Recall number
Z-2315-2018

Terminated Class II Medical devices

Recall started
29 December 2017
Terminated
29 November 2023
Reason
During an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.
Distribution
United States
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0843-2018, Z-0842-2018, Z-0845-2018, Z-0844-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.