Topcon Medical Systems, Inc. recalls

3 recalls recorded since 2018, most recently on 17 September 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)1
Terminated or completed2
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
17 September 2024
Reason
The overlay of visual field test locations on a probability map, as presented in IMAGEnet 6 in the so-called "Hood Report", is not appropriate as it can imply a structure/function correlation for which clinical evidence is not provided
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0362-2025

Terminated Class II Medical devices

Recall started
18 December 2020
Terminated
19 October 2022
Reason
Harmony RS integrations with Topcon equipment, TRC NW-400 and the Signal Camera, allowed user input of non-unique patient IDs that caused multiple patient records and images to be combined under the first generated record
Distribution
AZ, MO, NC, NY, PA, TX
Initiated
Voluntary: Firm initiated
Recall number
Z-0952-2021

Terminated Class II Medical devices

Recall started
6 March 2019
Terminated
17 July 2019
Reason
When a patient is selected in the patient data and another patient is selected prior to the display of the primary patient's thumbnails is completed, then the primary patient's thumbnails are displayed with secondary patient's identification information such as name, patient's ID, etc. on the monitor screen of the PC used.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-1138-2019