Owings Mills, MD, United States
Tollos recalls
1 recall recorded since 2018, most recently on 24 October 2022. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 24 October 2022
- Reason
- Two customers reported four incidents in which the lifting strap broke while lifting a patient. If the IC strap fails during use, the patient may fall causing injury to them or the caregiver(s).
- Distribution
- Worldwide - US Nationwide distribution and the countries of Canada and United Arab Emirates.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0245-2023
- Infection Control (IC) Lifting Strap used with the following Cirrus and Pinnacle Ceiling Lifts: Cirrus 450, Cirrus 600, Cirrus 700, Cirrus 750, and Pinnacle 1000
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.