TISSIUM SA recalls

1 recall recorded since 2018, most recently on 15 June 2026. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
15 June 2026
Reason
Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.
Distribution
US Nationwide distribution in the states of CA, FL, IL, NJ, OH, PA.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2903-2026, Z-2906-2026, Z-2907-2026, Z-2904-2026, Z-2901-2026, Z-2902-2026, Z-2908-2026, Z-2905-2026