Zurich, Switzerland
Thoratec Switzerland GMBH recalls
3 recalls recorded since 2018, most recently on 22 May 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls3
Ongoing (FDA)1
Terminated or completed2
Class I (FDA)1
Ongoing Class II Medical devices
- Recall started
- 22 May 2024
- Reason
- Due to a component change, the console part of a circulatory support system may not conform to the IEC 61000-4-5 standard, so if subjected to a power surge above 1.8kV and up to 2.0 kV, the console may shut down completely without alarm, pumping would cease, and the patient would have to be switched to a backup circulatory support system.
- Distribution
- Worldwide - US Nationwide distribution included in the states of CA, PA, FL, IL, NY, AR, GA, WI, AZ, TN, TX, NC, MT, VA, MA, NJ, NE, WV, ME, OK, CO, OH, KC, WA, IN, CT, MI, OR, ID, MO, SC, KY, LA, NM, NV, VT, IA, MD and the countries of ALGERIA, AUSTRALIA, AUSTRIA, BANGLADESH, BULGARIA, CANADA, CAYM
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2215-2024
- 2nd Generation CentriMag Primary Console, REF: 201-90401, 201-90701, 201-90411, 201-90421, 201-30300, L201-90401, L201-90411, L201-90421, part of the CentriMag Circulatory Support System
Terminated Class I Medical devices
- Recall started
- 22 August 2019
- Terminated
- 26 April 2024
- Reason
- Reports have been received of the CentriMag Systems experiencing motor and pump issues resulting from electromagnetic interference (EMI).
- Distribution
- Distribution was nationwide. There was government distribution, but no military distribution. Foreign distribution was made to Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Finland, France, Hong Kong, India, Israel, Italy, Liechtenstein, Mexico, Netherlands, Portugal, Qatar
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0221-2020
- CentriMag Acute Circulatory Support System Motor, U.S. Model 102956 and OUS Model 201-10002, a component of the CentriMag Extracorporeal Blood Pumping System.
Terminated Class II Medical devices
- Recall started
- 5 September 2018
- Terminated
- 28 April 2021
- Reason
- Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.
- Distribution
- U.S.: OR, AZ, CA, ID, WA, WI, NY, MA, CT, NJ, ME, RI, IL, MI, TN, OH, IN, KY, IA, AL, FL, NC, GA, SC, VA, MD, DC, PA, WV, DE, MN, MT, NE, CO, NM, UT, AR, MO, MC, TX, OK, LA, MS, ND
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0103-2019
- Thoratec CentriMag Motor, Model: 102956, a component of the Thoratec CentriMag Blood Pumping System and PediMag Blood Pumping System.