Amersham, United Kingdom
THOR Photomedicine Ltd recalls
3 recalls recorded since 2018, most recently on 24 July 2025. 3 are listed as ongoing. Status and classification are as published by the FDA.
Recalls3
Ongoing (FDA)3
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 24 July 2025
- Reason
- The potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
- Distribution
- Worldwide distribution: US (nationwide): AK, AZ, CA, CO, CT, FL, GA, HI, ID, IN, MA, MD, NC, ND, NJ, NV, NY, OR, PA, TX, UT, VA, WA, WI, WV; and OUS (International): Australia, United Kingdom, Austria, Ireland, Italy, Lithuania, Netherlands, Poland, Serbia, Slovakia, Spain.
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-2419-2025, Z-2417-2025, Z-2418-2025, Z-2416-2025
- Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Model/Catalog Number: S2186 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum t
- Brand Name: THOR Product Name: NovoTHOR XL Model/Catalog Number: S2190 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide top
- Brand Name: THOR Product Name: NovoTHOR Gen 3.0 Reg PRX Model/Catalog Number: S2189 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum t
- Brand Name: THOR Product Name: NovoTHOR Model/Catalog Number: S2188 Software Version: N/A Product Description: The NovoTHOR and NovoTHOR XL are intended to emit energy in the visible and IR spectrum to provide topica
Ongoing Class II Medical devices
- Recall started
- 20 November 2023
- Reason
- One of the ball studs attaching a gas strut to the canopy may fail. There is potential risk of injury should a ball stud fail when someone is in the NovoTHOR and trying to open or close the canopy
- Distribution
- Domestic distribution to the states of Arizona California Colorado Connecticut Florida Georgia Hawaii Idaho Indiana Maryland New Jersey New York Oregon Pennsylvania Texas Washington Wisconsin. Foreign distribution to Australia, Canada, Italy, Malta, Norway, Switzerland, UAE, and UK.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0972-2024
- NovoTHOR Whole Body Light Pod, Gen 3.0 (1) Product Code S2188 - Reg (2) Product Code S2189 - Reg PRX (3) Product Code S2190 - XL
Ongoing Class II Medical devices
- Recall started
- 13 September 2022
- Reason
- Failure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.
- Distribution
- Worldwide Distribution: US (Nationwide): Arizona, California, Colorado, Connecticut, Florida, Hawaii, Idaho, Illinois, Iowa, Maryland, Michigan, Minnesota, Missouri, Nevada, New Hampshire, North Carolina, Ohio, Oregon, Texas, Utah, Virginia, West Virginia. OUS (foreign) to countries of: Australia,
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-0455-2024, Z-0453-2024, Z-0454-2024, Z-0456-2024
- NovoTHOR Gen 2.5 Regular whole body red light therapy bed, model S2183
- NovoTHOR Gen 2.0 whole body red light therapy bed, model S1176
- G041 POD, THOR NOVO GEN 2.0 XL whole body red light therapy bed, model S1177
- NovoTHOR Gen 2.5 XL whole body red light therapy bed, model S2184.
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.