Recall Record

Fremont, CA, United States

Think Surgical, Inc. recalls

3 recalls recorded since 2018, most recently on 8 May 2026. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)2
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
8 May 2026
Reason
Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.
Distribution
U.S. Nationwide distribution in the states of AZ, CA, DE, FL, IA, KS, LA, MI, MN, MO, MS, NC, NE, NJ, NV, NY, OH, OR, PA, TX, UT, VA, and WA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2716-2026

Ongoing Class II Medical devices

Recall started
12 February 2024
Reason
There is a galling effect (cold-weld) caused by friction between the array fixation pins and the TMINI array guide holes due to insufficient clearance between the outer diameter of the fixation pins and the inner diameter of the TMINI array guide holes. As a result, the array guide has the possibility of binding to the array fixation pins during drilling.
Distribution
US Nationwide distribution in the states of AL, OH, NV, TX, UT, CA, MO.
Initiated
Voluntary: Firm initiated
Recall number
Z-1318-2024

Terminated Class II Medical devices

Recall started
30 March 2018
Terminated
24 January 2020
Reason
There is a potential for the cutter head to detach from the cutter shaft.
Distribution
The devices were distributed to the following US states: NJ, NY, OH, and TX. The devices were also distributed to Korea.
Initiated
Voluntary: Firm initiated
Recall number
Z-2605-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.