Thermo Fisher Scientific (Asheville) LLC recalls

1 recall recorded since 2018, most recently on 2 April 2019. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
2 April 2019
Terminated
23 July 2020
Reason
Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
Distribution
CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1868-2019, Z-1867-2019, Z-1866-2019, Z-1865-2019, Z-1873-2019, Z-1870-2019, Z-1869-2019, Z-1872-2019, Z-1871-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.