Thermo Fisher Scientific (Asheville) LLC recalls
1 recall recorded since 2018, most recently on 2 April 2019. None are listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0
Terminated Class II Medical devices
- Recall started
- 2 April 2019
- Terminated
- 23 July 2020
- Reason
- Product may have been assembled incorrectly,the assembly error can cause the bath to malfunction and raise the temperature of the fluid in the reservoir above the temperature set point selected by the user.
- Distribution
- CA, GA, KY, PA Foreign: Canada, China, Germany, Japan, New Zealand, Australia
- Initiated
- Voluntary: Firm initiated
- Recall numbers
- Z-1868-2019, Z-1867-2019, Z-1866-2019, Z-1865-2019, Z-1873-2019, Z-1870-2019, Z-1869-2019, Z-1872-2019, Z-1871-2019
- Thermo Scientific General Purpose Dynamic Bath Precision 10L GP Bath Model Number: TSGP10
- Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 20L GP Bath Model Number: FSGPD20
- Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 10L GP Bath Model Number: FSGPD10
- Fisher Scientific Isotemp -General Purpose Water Bath: lsotemp 5L GP Bath Model Number: FSGPD05
- Thermo Scientific General Purpose Dynamic Bath Precision 5L GP Bath Model Number: TSGP05
- Thermo Scientific General Purpose Dynamic Bath Precision 2L-Shallow GP Bath Model Number: TSGP2S
- Thermo Scientific General Purpose Dynamic Bath Precision 20L GP Bath Model Number: TSGP20
- Thermo Scientific General Purpose Dynamic Bath Precision 28L GP Bath Model Number: TSGP28
- and 1 more product lines
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.