Theralase Inc. recalls

6 recalls recorded since 2018, most recently on 26 September 2019. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls6
Ongoing (FDA)0
Terminated or completed6
Class I (FDA)0

Terminated Class III Medical devices

Recall started
26 September 2019
Terminated
3 December 2020
Reason
Device was shipped without a Unique Device Identification (UDI) label.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, LA, MD, MN, NC, NJ, NM, NY, OR, WI and countries of AU, BH, CA, HK, JO, MX, PK, PH.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0606-2020, Z-0607-2020

Terminated Class III Medical devices

Recall started
26 September 2019
Terminated
3 December 2020
Reason
Power supply (Model Number: TLC-2003) ME90A2403F01 (24 VDC, 3.75 A, 90 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 4th edition specification) shipped versus Power supply (Model Number: TLC-2003) MENB1100A2403F01 (24 VDC, 4.2 A, 101 W, 1% peak to peak ripple and noise, which meets UL/EN/IEC-60601-1-2, 3rd edition specification)
Distribution
Worldwide distribution - US Nationwide and countries Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania.
Initiated
Voluntary: Firm initiated
Recall number
Z-1326-2020

Terminated Class III Medical devices

Recall started
26 September 2019
Terminated
3 December 2020
Reason
Power Pack (component) was shipped with an internal battery pack tested to UN-38.3, UL-2054 and IEC-62133 (2nd Edition) standards; however, not tested to IEC-60601-1-2 (3rd edition) standards
Distribution
U.S.A.; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
Initiated
Voluntary: Firm initiated
Recall number
Z-1348-2020

Terminated Class III Medical devices

Recall started
26 September 2019
Terminated
3 December 2020
Reason
Laser Probe (Model Number: TLC-2001) Electromagnetic Interference ( EMI ) in excess of IEC-60601-1-2 (4th edition) standard
Distribution
Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
Initiated
Voluntary: Firm initiated
Recall number
Z-1266-2020

Terminated Class III Medical devices

Recall started
26 September 2019
Terminated
4 June 2020
Reason
905 nm laser diodes may be below rated power output.
Distribution
USA; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
Initiated
Voluntary: Firm initiated
Recall number
Z-1354-2020

Terminated Class III Medical devices

Recall started
26 September 2019
Terminated
3 December 2020
Reason
660 nm laser diodes may be below rated power output.
Distribution
USA; Australia, Bahrain, Canada, Hong Kong Province of China, Jordan, Mexico, Pakistan, Philippines, Thailand, Romania
Initiated
Voluntary: Firm initiated
Recall number
Z-1353-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.