Recall Record

Wilsonville, OR, United States

The Seaberg Company Inc recalls

1 recall recorded since 2018, most recently on 2 May 2018. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)0
Terminated or completed1
Class I (FDA)0

Terminated Class II Medical devices

Recall started
2 May 2018
Terminated
1 July 2020
Reason
Based on internal testing, results indicated a possible failure of the stitches securing the buckle to the nylon belt could occur, posing a potential risk when used on a human patient to stop arterial blood flow. There is a remote chance that a serious injury and/or death could occur as a result of the failure of the device.
Distribution
USA: CA, MD, OH, FL, ID, CO, CT, OR, AZ, IL, ME, MO, NY, SC, MI, AR, MT, CT, WA, TX, KY, NC, NJ, PA OUS: Australia, Canada, China, Columbia, Costa Rica, Europe (EU), Hong Kong, Iceland, India, Iraq, Israel, Japan, Kenya, Korea, Lebanon, Malaysia, Russia, Saudi Arabia, Singapore, South Africa, Tha
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2226-2018, Z-2228-2018, Z-2227-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.