Whitland, N/A, United Kingdom

The Magstim Company Limited recalls

5 recalls recorded since 2018, most recently on 8 December 2020. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)1
Terminated or completed4
Class I (FDA)0

Terminated Class II Medical devices

Recall started
8 December 2020
Terminated
16 February 2023
Reason
A small number of Neurosign 100 Intraoperative Nerve Monitors may have been dispatched without being configured appropriately for the supply mains within the target Country. The specific deficiency includes: 1. Onboard voltage selector set to 240V (European Standard), versus 120V requirement for the United States and Canada. 2. Fuses installed in the Power Entry Module (PEM) 2xT315mAL (240V Standard), versus the 2xT630mAL requirement for a 120V mains supply.
Distribution
MN, NC, and NY
Initiated
Voluntary: Firm initiated
Recall number
Z-0869-2021

Terminated Class II Medical devices

Recall started
23 June 2020
Terminated
16 August 2024
Reason
Reports that manipulation of the cable which connects the pre-amplifier to the nerve monitor, and the cable which connects the pre-amplifier to the stimulator pod, may cause damage to the internal construction of the cables.
Distribution
Domestic: MN International: UK, Republic of Ireland, Spain, Netherlands, Canada. (Technomed Europe distributed to: Brazil, Spain, Belgium, Chile, Greece, Poland, New Zealand, Pakistan, Portugal and Thailand.
Initiated
Voluntary: Firm initiated
Recall number
Z-2884-2020

Terminated Class II Medical devices

Recall started
23 March 2020
Terminated
20 September 2024
Reason
It was identified that there is a potential for the Stim Interconnecting high Voltage Cable located at the rear of the Rapid2 and Horizon power supply unit to fail.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, CA, CT, TN, IL, MN, MT, NC, NJ, NY, PA, SC, TX, WA, and the countries of Cyprus, China, Iran, Saudi Arabia, UK, Italy, Malaysia, Spain, Taiwan, Vietnam, Australia, Japan, Turkey, France and Israel.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1965-2020, Z-1966-2020

Ongoing Class II Medical devices

Recall started
25 October 2018
Reason
A stray strand of stainless-steel cable may protrude from the heat shrink sleeving on the Balance Cable of the Ez Arm.
Distribution
CA, MA, MN, PA, TX, VA, WI, and WA
Initiated
Voluntary: Firm initiated
Recall number
Z-0836-2019

Terminated Class II Medical devices

Recall started
28 September 2017
Terminated
21 June 2022
Reason
Users manuals not supplied with 4800-00T US Rapid Therapy System
Distribution
US Nationwide Distribution in the states of AR, CA, GA, KS, MA, MD, MO, NC, NJ, NW, NY, TX and Hawaii
Initiated
Voluntary: Firm initiated
Recall number
Z-1352-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.