Terumo Medical Corporation recalls

5 recalls recorded since 2018, most recently on 15 June 2026. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls5
Ongoing (FDA)1
Terminated or completed4
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
15 June 2026
Reason
Product was packaged with the 0.025 guidewire instead of the 0.021 guidewire.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-2930-2026

Terminated Class I Medical devices

Recall started
26 April 2019
Terminated
22 September 2020
Reason
There is a potential for dislodgement of the tip from the outer diameter of the sheath resulting in a loss of the smooth transition from the surface of the tip to the outer surface of the expandable sheath.
Distribution
The products were distributed US nationwide. The products were distributed to the following foreign countries: Australia, Belgium, Brazil, Canada, Portugal, Hong Kong, Peru, Singapore.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1374-2019, Z-1376-2019, Z-1375-2019

Terminated Class II Medical devices

Recall started
8 February 2019
Terminated
8 May 2020
Reason
There is a potential for a loss of package integrity that may compromise the sterility of the product.
Distribution
The products were distributed to the following US states: FL, IN, KS, KY, MA, MO, NH, NV, PA, SC, TN, and TX.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2003-2019, Z-2005-2019, Z-1998-2019, Z-1999-2019, Z-2000-2019, Z-2004-2019, Z-2006-2019, Z-2002-2019, Z-2001-2019

Terminated Class II Medical devices

Recall started
9 April 2018
Terminated
19 April 2019
Reason
Two lots of product were released for distribution prior to the completion of all required validation and lot acceptance activities. Although sterile, products from these lots may not function as intended and therefore should not be used.
Distribution
Distributed to accounts in AL, CA, FL, IA, IN, MI, MO, NC, NY, SD, TN, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-2337-2018

Terminated Class II Medical devices

Recall started
12 September 2017
Terminated
31 August 2018
Reason
The tucking depth feature of this catheter introducer is marginally outside the upper specification.
Distribution
To be added.
Initiated
Voluntary: Firm initiated
Recall number
Z-0725-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.