Recall Record

Windermere, FL, United States

TAS Medical Inc recalls

1 recall recorded since 2018, most recently on 27 January 2025. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
27 January 2025
Reason
Tissue approximation straps (zip-ties) have broken before or during implantation so firm developed an interim zip-tie conditioning method submerging zip-ties in 104-degrees F sterile saline solution for 60 min prior to implantation to rehydrate the zip-tie straps. Broken zip-ties not observed during procedure could lead to future hernia recurrence, recurrent surgery to correct a large herniation.
Distribution
US Nationwide distribution in the states of LA, OK, TN.
Initiated
Voluntary: Firm initiated
Recall number
Z-1294-2025

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.