Sysmex America, Inc. recalls

6 recalls recorded since 2018, most recently on 15 December 2025. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls6
Ongoing (FDA)2
Terminated or completed4
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
15 December 2025
Reason
Under specific conditions, the lift's electromagnetic holding force may disengage unexpectedly, allowing unintended downward movement of the lift.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1136-2026

Ongoing Class II Medical devices

Recall started
6 January 2025
Reason
Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
Distribution
US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1113-2025

Terminated Class II Medical devices

Recall started
6 July 2020
Terminated
21 May 2021
Reason
Insufficient amount of antibody without an error message or alarm
Distribution
US Nationwide distribution including in the states of AL, AZ, MD, WA, FL, IL.
Initiated
Voluntary: Firm initiated
Recall number
Z-2687-2020

Terminated Class II Medical devices

Recall started
9 March 2020
Terminated
24 March 2021
Reason
Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.
Distribution
USA nationwide distribution in the states of AK, AR, AZ, CA, CO, CT, DE, FL, IL, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, TN, TZ, UT and WA..
Initiated
Voluntary: Firm initiated
Recall number
Z-1703-2020

Terminated Class II Medical devices

Recall started
21 January 2020
Terminated
30 January 2023
Reason
When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.
Distribution
USA: FL, VT, IL.
Initiated
Voluntary: Firm initiated
Recall number
Z-1717-2020

Terminated Class II Medical devices

Recall started
27 July 2018
Terminated
8 July 2019
Reason
The fluorescent platelet counts (PLT-F) and immature platelet fraction (IPF) values can be falsely low when using affected lots of Fluorocell PLT.
Distribution
Worldwide Distribution: US (Nationwide) including Puerto Rico; and countries of: Argentina, Brazil, Canada, Chile, Costa Rica, Ecuador, Honduras, and Nicaragua.
Initiated
Voluntary: Firm initiated
Recall number
Z-2953-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.