Recall Record

Toronto, N/A, Canada

Synaptive Medical Inc recalls

3 recalls recorded since 2018, most recently on 23 April 2021. None are listed as ongoing. Status and classification are as published by the FDA.

Recalls3
Ongoing (FDA)0
Terminated or completed3
Class I (FDA)1

Terminated Class II Medical devices

Recall started
23 April 2021
Terminated
12 January 2024
Reason
Due to visible burrs/metal filings affixed along the suction tube inner perimeter.
Distribution
Domestic Distribution: IN, SC, MD, OR, TX ,LA, MI, NY, NV, FL, VA, CA,WI, PA and NC. International Distribution: Australia, Canada, Belgium, Netherlands, Singapore, Switzerland, Thailand, UK, Vietnam
Initiated
Voluntary: Firm initiated
Recall number
Z-1955-2021

Terminated Class II Medical devices

Recall started
26 February 2020
Terminated
10 April 2024
Reason
A software defect negatively affects the quality of images taken during diagnostic imaging (mammograms) which may result in misdiagnosis.
Distribution
US
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1758-2020, Z-1759-2020

Terminated Class I Medical devices

Recall started
9 October 2018
Terminated
12 August 2021
Reason
This recall has been initiated due to a software defect found in the Guide System software when used with NICO BrainPath ports under certain circumstances. When one trajectory is set with a blue port and another trajectory is set with a gold port, the software defect is triggered when the user switches between these trajectories during a surgical procedure.
Distribution
Worldwide - US Nationwide Distribution and the country of Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-0470-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.