Tempe, AZ, United States

Stryker Sustainability Solutions recalls

22 recalls recorded since 2018, most recently on 19 August 2026. 12 are listed as ongoing. Status and classification are as published by the FDA.

Recalls22
Ongoing (FDA)12
Terminated or completed10
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
19 August 2026
Reason
Reusable access port system may crack or seperate during use.
Distribution
US Nationwide distribution including in the states of Alabama, Alaska, Arizona, Arkansas, California, Colorado, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Mis
Initiated
Voluntary: Firm initiated
Recall numbers
Z-3183-2026, Z-3187-2026, Z-3186-2026, Z-3185-2026, Z-3184-2026, Z-3182-2026

Ongoing Class II Medical devices

Recall started
18 August 2026
Reason
Diagnostic electrophysiology (EP) catheter may have a breached sterile barrier on the straight seal side of the primary pouch, which if used could lead to particulate embolization, infection, thrombus or ganuloma formation from particulate matter.
Distribution
US Nationwide distribution in the state of NJ.
Initiated
Voluntary: Firm initiated
Recall number
Z-0085-2027

Ongoing Class II Medical devices

Recall started
10 April 2026
Reason
Incomplete seals on sterile product
Distribution
US Nationwide and the countries of Israel and Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2177-2026, Z-2174-2026, Z-2191-2026, Z-2181-2026, Z-2187-2026, Z-2195-2026, Z-2176-2026, Z-2192-2026, Z-2198-2026, Z-2185-2026, Z-2183-2026, Z-2194-2026 and 14 more

Ongoing Class II Medical devices

Recall started
12 March 2026
Reason
Due to potential for tourniquet cuffs to detach from the bladder at the welding connection and result in the cuff not being able to maintain pressure.
Distribution
U.S.: AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NY, OHJ, OK, OR, PA SC, SD, TN, TX, VA, VT, and WA. O.U.S.: N/A
Initiated
Voluntary: Firm initiated
Recall number
Z-1804-2026

Ongoing Class II Medical devices

Recall started
24 September 2025
Reason
Due to manufacturing issue, reprocessed shears, may become stuck in the "on position" resulting in continuous activation.
Distribution
US Nationwide distribution and OUS (foreign) to country of: Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0490-2026

Ongoing Class II Medical devices

Recall started
2 July 2024
Reason
Disposable 5 lead cable and lead wire system package that should contain lead systems, lead telemetry systems and a dual connect cable, may in fact contain, different ECG Leads, which are not approved for reprocessing, but that may have been reprocessed, which may lead to devices not performing as intended.
Distribution
US Nationwide distribution in the states of FL, WY, AZ, CA, LA, TX, MI, OR, CO, MS, TN, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2689-2024

Ongoing Class II Medical devices

Recall started
31 May 2024
Reason
Mislabeling; tourniquet cuff packaging labeled as dual port single bladder but contains a tourniquet cuff single port single bladder.
Distribution
US Nationwide distribution in the states of CA, CO, IL, MI, OR, WI.
Initiated
Voluntary: Firm initiated
Recall number
Z-2405-2024

Ongoing Class II Medical devices

Recall started
10 May 2024
Reason
Diagnostic electrophysiology (EP) catheters do not meet testing requirements.
Distribution
International distribution to the country of Japan.
Initiated
Voluntary: Firm initiated
Recall number
Z-2268-2024

Ongoing Class II Medical devices

Recall started
21 June 2023
Reason
39-inche-wide inflatable air transfer mattress packages contain 34-Inch-wide mattresses, so there may be a decreased surface area available to distribute the patient's weight across, which may increase the effort exerted by staff for a lateral patient transfer, and there may be increased friction and shearing forces when the patient is moved resulting in skin tear injury.
Distribution
US: WI, ID, CA, FL, HI, MD, RI, AZ, MA, GA, IL
Initiated
Voluntary: Firm initiated
Recall number
Z-2394-2023

Ongoing Class II Medical devices

Recall started
14 April 2023
Reason
Through complaint surveillance, it was identified that reprocessed Deep vein thrombosis (DVT) Compression Sleeves may be mislabeled and/or mispackaged.
Distribution
US: WA OUS: None
Initiated
Voluntary: Firm initiated
Recall number
Z-1612-2023

Ongoing Class II Medical devices

Recall started
1 March 2023
Reason
Increased incidence of disposable tourniquet cuffs "not able to achieve desired pressure" and "not holding pressure" due to 1) Cuff fabric tearing during use, and 2) Pinched or unseated O-ring at the connection between the cuff, disposable fill line, and/or cuff hosing, which may lead to loss of vascular occlusion, which may lead to operative site blood loss, hemorrhage, or other complications.
Distribution
US: WI, MD, NY, PA, WA, CO, GA, FL, OH, NC, TX, NJ, OR, MT, CA, RI, SC, NE, MI, LA, OK, SD, AL, VA, MS, WV, AZ, KY, IL, AR, ME, ND, MO, KS, HI, NH, AK, TN, IA, MN, IN, NV, MA, ID, DE, NM, WY, DC, VT, CT
Initiated
Voluntary: Firm initiated
Recall number
Z-1315-2023

Terminated Class II Medical devices

Recall started
21 January 2022
Terminated
26 February 2025
Reason
Mislabeling; dual-bladder tourniquet cuffs are labeled as single-bladder inflatable cuffs.
Distribution
US: KS OK TX OUS: None
Initiated
Voluntary: Firm initiated
Recall number
Z-0664-2022

Ongoing Class II Medical devices

Recall started
19 January 2021
Reason
Product was distributed without receiving regulatory clearance.
Distribution
Worldwide distribution - U.S. Nationwide distribution in the state of OH and the country of Israel.
Initiated
Voluntary: Firm initiated
Recall number
Z-1327-2021

Terminated Class II Medical devices

Recall started
27 March 2020
Terminated
14 October 2022
Reason
A subset of temporary intracardiac ultrasound catheters are labeled with the incorrect expiration date. They are mislabeled as having a shelf life of two years, while the actual shelf life is one year. The earliest affected product will expire in May 2020. Use of mislabeled devices may lead to hazards sterility breach, component breakdown or delay in treatment.
Distribution
US: AZ, CA, FL, GA, ID, IL, IN, KS, MA, MO, PA, TN, TX, VA, WA, WV OUS: None
Initiated
Voluntary: Firm initiated
Recall number
Z-2113-2020

Terminated Class II Medical devices

Recall started
9 October 2019
Terminated
14 January 2021
Reason
Tourniquet cuffs are potentially mislabeled. If used may lead to (1) Device misuse during procedure (2) Inadequate Limb Compression Force.
Distribution
US Nationwide distribution including the states of SC, TX, UT, NY, AR
Initiated
Voluntary: Firm initiated
Recall number
Z-1802-2020

Terminated Class II Medical devices

Recall started
17 December 2018
Terminated
8 December 2023
Reason
Stryker s Sustainability Solutions division (SSS) has received an increase in reports indicating that an EEPROM chip error code may occur when Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are used with CARTO(R) EP Navigation Systems.
Distribution
US distribution to states of: GA, KS, MO FL, CA, WA, IA,; and internationally to: Germany, Japan and Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0766-2019

Terminated Class II Medical devices

Recall started
23 August 2018
Terminated
4 February 2021
Reason
Electrophysiology catheters were identified to be item # D134903 and distributed. It was subsequently determined that these catheters are item # D134909, which is not approved for reprocessing.
Distribution
US: KS OUS: None
Initiated
Voluntary: Firm initiated
Recall number
Z-0467-2019

Terminated Class II Medical devices

Recall started
27 July 2018
Terminated
17 September 2020
Reason
Affected lots of reprocessed arthroscopic shavers may display the incorrect catalog number and speed settings when connected to the generator, require manual mode for use on CORE generators, or display an error message on Crossfire 1 and 2 generators which does not allow the device to be used.
Distribution
Nationwide Distribution: AL, AR, AZ, CA, CO, DC, DE, FL, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY and Puerto Rico. OUS Israel and Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0973-2019

Terminated Class II Medical devices

Recall started
19 June 2018
Terminated
15 November 2024
Reason
Through investigation, it was determined that H12LP Trocars in scope of a previous H12LP recall were distributed to facilities within the distribution chain following the recall execution activities initially performed by distributor.
Distribution
US nationwide distribution in the states of OK, GA, NJ, TX, IA, CO, TX, CA, LA, PA, VA, MD, WA, and IL. OUS: None.
Initiated
Voluntary: Firm initiated
Recall number
Z-0498-2019

Terminated Class II Medical devices

Recall started
8 May 2018
Terminated
29 July 2019
Reason
Stryker Sustainability Solutions has received an increase in reports indicating that the affected Reprocessed HARH36 devices may display an error code (No Instrument Uses Remaining) upon initial connection to the generator. When this error code is present, the device is not able to be used.
Distribution
US Nationwide Distribution in the states of PA, FL, MD, MA, CA, NY, NC, TX, MT, AL, LA, TN
Initiated
Voluntary: Firm initiated
Recall number
Z-2484-2018

Terminated Class II Medical devices

Recall started
29 March 2018
Terminated
28 July 2020
Reason
Thrombogenicity test results demonstrated that both the test devices (SSS) and control devices (OM) did not meet the internal specification of non-thrombogenic .
Distribution
Worldwide Distribution - US Nationwide and Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-2426-2018

Terminated Class II Medical devices

Recall started
2 January 2018
Terminated
2 May 2018
Reason
An EEPROM chip error code may occur when the catheters are used with CARTO EP Navigation Systems.
Distribution
USA (nationwide) Distribution was made to medical facilities in AZ, CA, CT, FL, GA, IL, KS, LA, MN, MO, OH, PA, TX, VT, WA, and Wi. Foreign distribution was made to Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0916-2018, Z-0917-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.