Fremont, CA, United States

Stryker Neurovascular recalls

15 recalls recorded since 2018, most recently on 6 April 2026. 6 are listed as ongoing. Status and classification are as published by the FDA.

Recalls15
Ongoing (FDA)6
Terminated or completed9
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
6 April 2026
Reason
Embolization coil detachment system devices may experience premature battery drain causing devices to: 1) Not power on; 2) Power on with faint audible and visual indicators; 3) Be unable to detach a coil as intended, which may necessitate medical intervention if the procedure must be completed with an alternative technique.
Distribution
Worldwide - US Nationwide distribution including in the states of PA, AZ, FL, CA, IL, AK, NY, MI, NJ, WI, SD, KY, AR, TX, AL, TN, MO, MA, IN, NE, WV, MT, CT, NC, VA, NV, CO, DC, DE, OH, ME, GA, WA, SC, GU, MN, OK, NH, MD, ND, UT, HI, OR, LA, MS, KS, NM, ID, IA, VT and the countries of ARGENTINA, AUS
Initiated
Voluntary: Firm initiated
Recall number
Z-2552-2026

Ongoing Class II Medical devices

Recall started
30 August 2024
Reason
Stryker Neurovascular has observed that devices with Out of United States (OUS) approved Instructions for Use (IFU) and Packaging Labels (pouch and carton), different configurations, have been erroneously distributed to customers in the US.
Distribution
U.S. Nationwide distribution in the states of CA and VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-0021-2025

Ongoing Class II Medical devices

Recall started
23 April 2024
Reason
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
Distribution
Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Aus
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1903-2024, Z-1906-2024, Z-1909-2024, Z-1908-2024, Z-1905-2024, Z-1902-2024, Z-1907-2024, Z-1904-2024

Ongoing Class II Medical devices

Recall started
18 April 2024
Reason
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
Distribution
Worldwide - US Nationwide distribution including in the states of TX, NY, MD, CA, KY, WI, OR, NE, SC, MA, MO, CT, ID, PA, TN, FL, NC, AL, IA, OK, VA, IL, OH, NH, MN, ND, KS, NJ, WV, IN, GA, HI, AR, UT, LA, MI, WA, NV, CO, AZ, DC, VT, ME, DE, AK, WY, MT, NM, SD, MS, RI, GU and the countries of ARGENT
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1941-2024, Z-1940-2024, Z-1939-2024

Ongoing Class II Medical devices

Recall started
3 April 2024
Reason
Stryker Neurovascular is recalling their DAC- Distal Access Catheter -038 136cm, a Distal Access Catheter by removal. The reason for the recall is: the DAC Lot #0000486382 was released with out-of-specification endotoxin results.
Distribution
US Nationwide distribution in the states of TX, FL, NJ, CO, LA, NC, AL, PA, MI, MS, SC, IL.
Initiated
Voluntary: Firm initiated
Recall number
Z-1740-2024

Ongoing Class II Medical devices

Recall started
3 June 2022
Reason
Microcatheter product label on the carton is missing the microcatheter product drawing/illustration.
Distribution
Worldwide distribution - US Nationwide distribution in the states of CA, OR, PA, NY, DC, IL, WA, NH, NJ, OK, FL, AZ, AL, GA, TX, WI, VA and the countries of DE, IT, AT.
Initiated
Voluntary: Firm initiated
Recall number
Z-1386-2022

Terminated Class I Medical devices

Recall started
21 September 2020
Terminated
1 April 2024
Reason
Increase in fracture complaints of the flexible, tapered core wire, resulting in stent retriever separation from the core wire during use. The stent retriever can be left behind in the vasculature.
Distribution
U.S.: PA, OH, AK, NY, VA, IL, CA, NJ, AL, NV, FL, KY, AR, MA, TX, MI, NC, TN, WI, WA, AZ, OR, GA, VT, UT, DC, MO, IN, OK, NH, LA, HI, DE, SC, MS, RI, MD, SD, KS, ID, CO, CT, IA, NM, WV, MN, GU, NE. Foreign: New Zealand, Argentina, Estonia, Lithuania, Slovenia, Australia, Finland, Luxembourg, Sout
Initiated
Voluntary: Firm initiated
Recall number
Z-0114-2021

Terminated Class II Medical devices

Recall started
18 April 2019
Terminated
27 April 2021
Reason
A design related issue can contribute to the device exhibiting brittle behavior. This behavior may result in fractures of the catheter shaft.
Distribution
U.S Nationwide Distribution - PA, VA, TX, CA, FL, AR, TN, MI, NY, IL, NC, KY, WV, OH, MO, AZ, GA, WA, VT, DC, SC, NJ, MN, OK, MD, AL, LA, MA, DE, IN, OR, RI, SD, MS, WI, ID, UT, CO, IA, KS, CT, NE, NM
Initiated
Voluntary: Firm initiated
Recall number
Z-1708-2019

Terminated Class II Medical devices

Recall started
10 April 2019
Terminated
25 February 2021
Reason
The firm has become aware of a potential problem that their Catheters may break/fracture and/or kink/bend if improperly removed from the packaging/dispenser hoop or during catheter use.
Distribution
US Nationwide distribution in the states of PA, FL, CA, IL, WI, AZ, AR, FL, TN, MI, MA, CA, NC, WV, DE, OH, NJ, MO, NY, SC, TX, AL, VA, UT, MD, GA, LA, CO, WA, KY, OK, RI, SD, MS, NE,IA, KS, DC. No US Govt. No Foreign countries.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1620-2020, Z-1621-2020

Terminated Class II Medical devices

Recall started
13 November 2018
Terminated
3 August 2020
Reason
Labeling error; 8F Balloon Guide Catheters have been packaged within 9F Balloon Guide Catheter packaging
Distribution
OUS distribution only: CZECH REPUBLIC, ITALIA, Netherlands, Sweden, Germany, Austria, Israel, France, Belgium, Finland, Switzerland, Portugal, Finland
Initiated
Voluntary: Firm initiated
Recall number
Z-0065-2020

Terminated Class II Medical devices

Recall started
8 June 2018
Terminated
28 April 2021
Reason
The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.
Distribution
Worldwide distributions- U.S. Nationwide including the states of AL, AZ, AR, CA, CO, CT, DC, DE, FL, GA, GU, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY. Countries of Austria, Argent
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0225-2020, Z-0228-2020, Z-0229-2020, Z-0232-2020, Z-0231-2020, Z-0230-2020, Z-0227-2020, Z-0226-2020

Terminated Class II Medical devices

Recall started
5 June 2018
Terminated
30 November 2020
Reason
Use-related coating damage on delivery catheter shafts of Wingspan delivery catheters were reported
Distribution
China
Initiated
Voluntary: Firm initiated
Recall number
Z-3020-2018

Terminated Class II Medical devices

Recall started
2 May 2018
Reason
The AXS Infinity LS Long Sheath Dilator outer diameter (OD) specification was exceeded.
Distribution
Only U.S. Consignees: CO,NJ,NY, OH,PA, and WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-3281-2018

Terminated Class II Medical devices

Recall started
11 April 2018
Terminated
28 April 2021
Reason
There may be product manufactured with the outer diameter of the balloon catheter exceeding the defined specification at a customer location. If the user experiences resistance between the FlowGate and an interfacing device, the device would be removed and a second device would be used, which could potentially prolong a procedure.
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-2569-2018

Terminated Class II Medical devices

Recall started
28 November 2017
Terminated
28 April 2020
Reason
Stryker Neurovascular has become aware that at least 1 (one) Trevo XP Provue System was manufactured where the carton sleeve did not match the physical contents of the pack.
Distribution
OUS only - NL, FR, IT, DE, BE
Initiated
Voluntary: Firm initiated
Recall number
Z-0347-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.