Stryker Medical Division of Stryker Corporation recalls

18 recalls recorded since 2018, most recently on 18 February 2026. 7 are listed as ongoing. Status and classification are as published by the FDA.

Recalls18
Ongoing (FDA)7
Terminated or completed11
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
18 February 2026
Reason
Cribs sold in the USA are missing two access door warning labels.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1580-2026

Ongoing Class II Medical devices

Recall started
31 October 2025
Reason
Stryker has issued a voluntary recall of certain MV3 beds and Arise 1000EX mattresses because some of these products have been sold in close proximity, suggesting a potential intent for them to be used together. These products have not been tested for compatibility and thus are not designed to be used together. The combination of both is not recommended by Stryker.
Distribution
US, nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0919-2026, Z-0918-2026

Ongoing Class II Medical devices

Recall started
30 August 2023
Reason
Patient beds are missing test values for electrical safety tests presenting risk of tissue burn or electrical shock.
Distribution
Distribution US Nationwide including Puerto Rico, Australia, Canada, Chile, China, Mexico, Poland, Spain and Switzerland.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0427-2024, Z-0425-2024, Z-0432-2024, Z-0435-2024, Z-0429-2024, Z-0436-2024, Z-0428-2024, Z-0434-2024, Z-0431-2024, Z-0430-2024, Z-0433-2024, Z-0424-2024 and 1 more

Ongoing Class II Medical devices

Recall started
24 May 2023
Reason
Base leg assembly may bend and result in difficulty engaging/disengaging from fastener, difficulty loading/unloading from ambulance, difficulty raising or lowering the cot, cot tip, or erratic or unintended motion.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Canada, Netherlands, Switzerland, United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-2130-2023

Ongoing Class II Medical devices

Recall started
1 February 2023
Reason
Transport Cot has insufficient volume of epoxy to assemblies installed in the head section and foot sectioning system may result in separation of head end or foot end from the cot and result in unexpected disengagement while loading/unloading or transport.
Distribution
US Nationwide distribution in the states of NY, VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1256-2023

Ongoing Class II Medical devices

Recall started
11 November 2022
Reason
Cot may experience unintended motion due to damaged cables. Potential risk of soft tissue injury to patients such as scrapes or bruises resulting from the cot lowering faster than expected.
Distribution
Worldwide Distribution: US (Nationwide) and Foreign to countries of: Australia, Canada, Netherlands, and United Kingdom.
Initiated
Voluntary: Firm initiated
Recall number
Z-0479-2023

Ongoing Class II Medical devices

Recall started
12 May 2021
Reason
Medical devices intended for use in a healthcare facility were distributed to residential customers and may be in use in a home environment. The product s labeling accurately identifies the correct use environment, but home users may not be aware of or fully understand these instructions intended for healthcare professionals
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Argentina, Brazil, Canada, Chile, Colombia, India, Mexico, Thailand.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2207-2021, Z-2211-2021, Z-2208-2021, Z-2216-2021, Z-2219-2021, Z-2210-2021, Z-2214-2021, Z-2205-2021, Z-2221-2021, Z-2220-2021, Z-2212-2021, Z-2218-2021 and 7 more

Terminated Class II Medical devices

Recall started
12 August 2020
Terminated
2 April 2021
Reason
Connector between support surface and pump may be out of tolerance resulting in difficulty in removing the CPR connector causing a delay in resuscitation efforts for patients experiencing cardiac arrest and needing CPR
Distribution
US Nationwide distribution including in the states of CA, FL, NY, OH.
Initiated
Voluntary: Firm initiated
Recall number
Z-3015-2020

Terminated Class II Medical devices

Recall started
7 April 2020
Terminated
22 January 2021
Reason
Ambulance Cots may not meet crash test standard BS EN 1789
Distribution
Foreign: UK
Initiated
Voluntary: Firm initiated
Recall number
Z-1935-2020

Terminated Class II Medical devices

Recall started
20 November 2018
Terminated
9 July 2020
Reason
The power load floor plates which hold the ambulance cot anchors in place during patient transport, may be broken.
Distribution
US Nationwide and the following foreign countries: Australia, Canada, Netherlands, South Africa, and Switzerland.
Initiated
Voluntary: Firm initiated
Recall number
Z-1350-2019

Terminated Class II Medical devices

Recall started
10 October 2018
Terminated
29 April 2020
Reason
It was identified that the volume of air inside the mattress may expand in regions of high altitude which may result in an unintended gap between the plateau of the inflated patient surface and tub. A Health Hazard Evaluation was completed which identified the potential risk of patients moving out of position and rolling over on the mattress. While the highest possible severity of harm associated with this condition is death, there have been no reports of injury or harm. This condition has only presented in regions of high altitude; however, Stryker has chosen to replace all Naera mattresses in the field.
Distribution
Nationwide distribution. Foreign distribution to Argentina, Australia, Canada, Chile, China, Colombia, Germany, India, Japan, Netherlands, South Africa, Turkey, Saudi Arabia, United Kingdom, and Mexico.
Initiated
Voluntary: Firm initiated
Recall number
Z-0813-2019

Terminated Class II Medical devices

Recall started
8 October 2018
Terminated
28 April 2020
Reason
The beds with the high rail side rail option may not be complaint with IEC 60601-2-52/C1:2010. There is a potential risk of entrapment.
Distribution
The products were distributed to the following US states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WY. The products were distributed to t
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0438-2019, Z-0439-2019

Terminated Class II Medical devices

Recall started
6 September 2018
Terminated
13 April 2020
Reason
Users may experience alarm fatigue due to frequent alarming relating to patient temperature deviation beyond 0.5¿C while in Automatic mode. 2. Users may experience a Remove from Use code 9 (RFU 9) fault condition relating to inadvertent flow alarms.3. Users may experience code RFU 27 fault condition relating to mechanical interference between the device fan and filter
Distribution
Worldwide - US Nationwide Distribution and the countries of Foreign: Australia, Canada, Netherlands, Switzerland and United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-0235-2019

Terminated Class II Medical devices

Recall started
13 August 2018
Terminated
21 May 2020
Reason
Certain 2860 IsoFlex LAL support surfaces were assembled with a non-conforming fire barrier.
Distribution
US Nationwide in the states of: KY, NH, OH, PA, and TX.
Initiated
Voluntary: Firm initiated
Recall number
Z-3181-2018

Terminated Class II Medical devices

Recall started
12 March 2018
Terminated
12 May 2020
Reason
The power load floor plates which hold the ambulance cot anchors in place during patient transport, may be broken.
Distribution
Worldwide Distribution - USA (nationwide) and to the countries of : Australia, Canada, France, Germany, Hong Kong, Italy, India, Netherlands, Portugal, Saint Martin, Switzerland, United Kingdom
Initiated
Voluntary: Firm initiated
Recall number
Z-1389-2018

Terminated Class II Medical devices

Recall started
9 March 2018
Terminated
9 April 2019
Reason
The Operations Manual and Maintenance Manual are missing various warnings and contain incorrect information.
Distribution
Product distributed to the Netherlands
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1599-2018, Z-1598-2018

Terminated Class III Medical devices

Recall started
7 November 2017
Terminated
9 April 2019
Reason
The T/Pump Operations Manual and Maintenance Manual were incorrectly revised to add the sentence, Temperature therapy can also be for heat conservation for normothermia maintenance to the Indications for Use statement.
Distribution
Distributed in all 50 states, the District of Columbia, and Guam.
Initiated
Voluntary: Firm initiated
Recall number
Z-1013-2018

Terminated Class II Medical devices

Recall started
21 July 2017
Terminated
22 March 2018
Reason
It was identified that the Mistral-Air Premium blankets may be capable of conducting electricity during a defibrillation or cardioversion therapy at higher energy levels (70 Joules or more) when the blanket is applied with the reflective side toward the patient (upside down).
Distribution
U.S.
Initiated
Voluntary: Firm initiated
Recall number
Z-0419-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.