Stryker Leibinger GmbH & Co. KG recalls

6 recalls recorded since 2018, most recently on 4 April 2024. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls6
Ongoing (FDA)3
Terminated or completed3
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
4 April 2024
Reason
Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-1885-2024

Ongoing Class II Medical devices

Recall started
29 March 2024
Reason
Mislabeled: Product labeled 56-90614 (90x90 Gold Mesh), was actually 56-90314 (90x90 Blue Mesh), may cause a delay in procedure
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall number
Z-1785-2024

Ongoing Class II Medical devices

Recall started
6 December 2022
Reason
Device was shipped without having regulatory approval.
Distribution
Distribution of one unit only to a single customer in MA (USA)
Initiated
Voluntary: Firm initiated
Recall number
Z-1234-2023

Terminated Class II Medical devices

Recall started
29 December 2021
Terminated
21 December 2023
Reason
Inside the packaging of one reported item a non-conforming product was found. Contrary to the design, the barrier layer was manufactured on the inferior side of the implant versus the superior side.
Distribution
Worldwide distribution - US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-0715-2022

Terminated Class II Medical devices

Recall started
1 April 2021
Terminated
19 December 2022
Reason
Inability for the user to inject the paste from the syringe into the target location.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Germany, Italy, France, Australia, Japan.
Initiated
Voluntary: Firm initiated
Recall number
Z-1436-2021

Terminated Class II Medical devices

Recall started
22 June 2018
Terminated
8 July 2019
Reason
After a three year term, there is a potential for the external packaging (sterile barrier) to become compromised.
Distribution
US distribution in the states: AZ, FL, NH, and OH, Australia, Canada, France, New Zealand, and UK.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2910-2018, Z-2902-2018, Z-2901-2018, Z-2912-2018, Z-2899-2018, Z-2915-2018, Z-2900-2018, Z-2911-2018, Z-2913-2018, Z-2907-2018, Z-2903-2018, Z-2914-2018 and 6 more

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.