Steris Corporation recalls

33 recalls recorded since 2018, most recently on 23 September 2026. 15 are listed as ongoing. Status and classification are as published by the FDA.

Recalls33
Ongoing (FDA)15
Terminated or completed18
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
23 September 2026
Reason
A software issue may result in unintended mismatch of patient information on manually printed images. Incorrect patient information appearing on printed images could create a documentation discrepancy or incorrect association of clinical images with a patient record.
Distribution
Worldwide - US Nationwide and the countries of Australia, Canada, New Zealand.
Initiated
Voluntary: Firm initiated
Recall number
Z-0118-2027

Ongoing Class II Medical devices

Recall started
21 September 2026
Reason
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0154-2027, Z-0156-2027, Z-0155-2027

Ongoing Class II Medical devices

Recall started
31 August 2026
Reason
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0074-2027, Z-0070-2027, Z-0073-2027, Z-0072-2027, Z-0071-2027, Z-0075-2027

Ongoing Class II Medical devices

Recall started
15 May 2026
Reason
The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2449-2026, Z-2450-2026

Ongoing Class II Medical devices

Recall started
15 May 2026
Reason
There is a potential for paint delamination (chipping/peeling) on the lighting system's light heads.
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Canada, India, New Zealand, Panama.
Initiated
Voluntary: Firm initiated
Recall number
Z-2448-2026

Ongoing Class II Medical devices

Recall started
19 November 2025
Reason
Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
Distribution
Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0909-2026, Z-0910-2026

Ongoing Class II Medical devices

Recall started
23 June 2025
Reason
The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
Distribution
Worldwide - US Nationwide and Puerto Rico and the country of Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2178-2025, Z-2177-2025

Ongoing Class II Medical devices

Recall started
23 May 2025
Reason
This is a sub-recall of event RES 96885. The supplier identified that during their manufacturing process for a limited batch of MNA-6x0 (adapter) components the incorrect configuration file was used.
Distribution
US Nationwide distribution in the states of FL, NC, NY, OH, TN, TX, and VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1361-2026

Ongoing Class II Medical devices

Recall started
13 June 2024
Reason
IFU for hookup were incorrectly reflected within hookup IFUs and online reference system, Hookup Lookup (HULU). Use of an incorrect hookup connection and/or parameter set may prevent an endoscope from being properly disinfected, which may present a risk of patient contamination or infection
Distribution
Worldwide distribution - United States Nationwide and the countries of Canada, Belgium, Czech Republic, Denmark, Germany, Spain, Italy, Slovenia, Slovakia, Australia, Israel, Jordan, Kuwait, Namibia, Saudi Arabia, South Africa, Cote d Ivoire, Afghanistan, Albania, Belarus, Botswana, Kazakhstan, Mala
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2538-2024, Z-2536-2024, Z-2537-2024, Z-2539-2024

Ongoing Class II Medical devices

Recall started
12 December 2023
Reason
Dual-articulating head rest may not stay in place or may fail to lock into position during a patient procedure which could result in an injury to the patient or a procedural delay.
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, FL, IL, IN, IA, KS, LA, MN, MS, NV, NJ, OH, OK, OR, PA, TN, TX, UT, WA, WI & Puerto Rico, and the countries of Australia, Canada, Mexico & New Zealand.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0983-2024, Z-0984-2024

Ongoing Class II Medical devices

Recall started
16 November 2023
Reason
Bandage Scissors handle cover component contains an incorrect color masterbatch resin carrier. This may allow a pink dye/residue to appear on the Bandage Scissors and/or sterilization pouches after processing.
Distribution
US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall number
Z-0737-2024

Ongoing Class II Medical devices

Recall started
13 November 2023
Reason
Identification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement of patient procedures.
Distribution
Domestic distribution nationwide. Foreign distribution to Australia Canada China Hong Kong Italy Malaysia Netherlands New Zealand Singapore Thailand
Initiated
Voluntary: Firm initiated
Recall number
Z-0617-2024

Ongoing Class II Medical devices

Recall started
10 April 2023
Reason
In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.
Distribution
Worldwide Distribution: US (Nationwide) and OUS (Foreign) to countries of: China, Lithuania, Saudi Arabia, Korea, Republic of, Canada, Taiwan, United Kingdom, Spain, Australia, New Zealand, Spain, Brazil, Bahrain, Switzerland, Great Britain, South Africa, Hong Kong, Qatar.
Initiated
Voluntary: Firm initiated
Recall number
Z-1595-2023

Ongoing Class II Medical devices

Recall started
17 March 2023
Reason
Light handle covers may separate from the light handle (detach and fall off) during use, this could result in a procedure delay and/or impact the sterile field which could result in an injury to the patient
Distribution
Worldwide - US Nationwide distribution including Puerto Rico and the countries of Canada, Australia.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1475-2023, Z-1476-2023

Ongoing Class II Medical devices

Recall started
12 December 2022
Reason
Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.
Distribution
Nationwide Foreign: AE AT CA CO ES FI GU IN IT JP MX SG TH TW ZA
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0781-2023, Z-0779-2023, Z-0780-2023, Z-0778-2023

Terminated Class II Medical devices

Recall started
17 October 2022
Terminated
28 May 2024
Reason
Two electronic diode components (D1 & D4) placed in the reverse orientation on the controller printed circuit board assembly may result in premature wear of the power regulator, leading to the regulators failure which could cause a full power loss. If the regulator failure were to occur during a patient procedure, a procedural delay may result
Distribution
US Nationwide distribution in the states of Arizona, Illinois, Tennessee, Virginia.
Initiated
Voluntary: Firm initiated
Recall number
Z-0241-2023

Terminated Class II Medical devices

Recall started
2 September 2022
Terminated
2 January 2025
Reason
The scalpel holder may separate from the handle for certain lots. If the tip falls off during use, it could fall into the patient's mouth which may result in a cut to the patient's mouth and also presents a choking hazard.
Distribution
Distribution to US states of IL, MA, NC, SC, and TN, Canada, China, Japan, Malaysia, Turkey, Great Britain, and India.
Initiated
Voluntary: Firm initiated
Recall number
Z-0079-2023

Terminated Class II Medical devices

Recall started
22 April 2022
Terminated
4 November 2024
Reason
The firm identified that when the electrical contactor component in the drying chamber of the Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.
Distribution
Worldwide distribution - US Nationwide.
Initiated
Voluntary: Firm initiated
Recall number
Z-1256-2022

Terminated Class II Medical devices

Recall started
22 March 2022
Terminated
29 January 2024
Reason
Product not approved for release for US distribution
Distribution
Nationwide
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1118-2022, Z-1119-2022

Terminated Class II Medical devices

Recall started
21 January 2021
Terminated
10 November 2021
Reason
On January 7, 2021, STERIS identified that the glued bond between the beam and the leg spar assembly may begin to separate over time. This could result in unexpected movement of the affected leg spar if the separation occurred during a patient procedure.
Distribution
Domestic: AL, AK, AR, AZ, CA, CO, CT, DC, FL, GA, ID, IL IN, IW, KS, KY, LA, ME, MD, MI, MN, MO, NE, NV, NJ, NY, NC, ND, OH, OK, OR, PA, PR, SC, TX, UT, VT, VA, WA, WV, WI. Foreign: Canada, South Korea, Panama, Qatar, Thailand.
Initiated
Voluntary: Firm initiated
Recall number
Z-1149-2021

Terminated Class II Medical devices

Recall started
11 December 2020
Terminated
21 June 2022
Reason
The electrical contactor component present in the drying chamber of the Reliance Vision Single-Chamber Washer/Disinfector may malfunction, potentially leading to the heating elements in the drying chamber to overheat, eventually resulting in smoke and/or fire.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Bulgaria, Canada, Chile, China, Ecuador, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Korea, Republic of Malaysia, Mexico, New Zealand, Panama, Portugal, Russian Federation, Saudi A
Initiated
Voluntary: Firm initiated
Recall number
Z-0876-2021

Terminated Class II Medical devices

Recall started
2 December 2020
Terminated
21 June 2022
Reason
Purple ink located on the outside label of Vis-U-All Pouches may migrate ( bleed ) following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch
Distribution
Nationwide Foreign:
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0792-2021, Z-0789-2021, Z-0794-2021, Z-0793-2021, Z-0791-2021, Z-0788-2021, Z-0796-2021, Z-0795-2021, Z-0797-2021, Z-0799-2021, Z-0798-2021, Z-0790-2021 and 2 more

Terminated Class II Medical devices

Recall started
19 November 2020
Terminated
25 October 2023
Reason
STERIS has identified that certain lots of diluted 4-Zyme may contain bacteria, specifically Pseudomonas fluorescens. The presence of this bacteria can cause the color of the detergent to darken over time. There is an improbable risk to users of the product from exposure to this bacteria, and no risk to patients.
Distribution
Worldwide distribution - US Nationwide distribution including in the states of CA, DE, CO, VA, FL, WA, TX, OH, PA, MA, IA, CT and the countries of United Kingdom, Saudi Arabia, Portugal, Denmark, Ireland, United Arab Emirates, South Korea, Australia, Greece, Ashmore & Cartier Islands, Belgium, Cana
Initiated
Voluntary: Firm initiated
Recall number
Z-0830-2021

Terminated Class III Medical devices

Recall started
7 October 2020
Terminated
14 July 2021
Reason
Potential for cleaning solution intrusion into endoscope
Distribution
Nationwide Foreign: Canada, Italy, Kuwait, United Kingdom, United Arab Emirates
Initiated
Voluntary: Firm initiated
Recall number
Z-0458-2021

Terminated Class II Medical devices

Recall started
26 June 2020
Terminated
22 December 2020
Reason
A recent FDA inspection at the firm identified that the foil test performed during installation of the device does not have clear acceptance criteria.
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Canada, Mexico, and United Arab Emirates.
Initiated
Voluntary: Firm initiated
Recall number
Z-2700-2020

Terminated Class II Medical devices

Recall started
22 October 2019
Terminated
9 June 2020
Reason
The locking pin of the front wheel asse mbly of the Evolution Transfer Carriages may not remain full y engaged with the transfer carriage and result in the transfer carriage and loading car unexpectedly falling forward
Distribution
Nationwide Foreign: Canada, Peru, Mexico, Brazil, Saudi Arabia, Columbia
Initiated
Voluntary: Firm initiated
Recall number
Z-0586-2020

Terminated Class II Medical devices

Recall started
22 October 2019
Terminated
6 May 2020
Reason
Potential for the incorrect pressure safety valve being included in the Preventive Maintenance (PM) kit for the 3000 Small Sterilizers.
Distribution
Worldwide distribution - US Nationwide distribution in the states of AL AR AZ CA CO DE FL GA HI IA ID IL IN KS LA MD ME MI MN MO MS MT NC NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA WA WI WV, and country of Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-0604-2020

Terminated Class II Medical devices

Recall started
26 December 2018
Terminated
16 June 2020
Reason
The ATLAS loading car may not remain fully engaged with the latch on the transfer carriage when the loading car is empty, potentially resulting in the loading car unexpectedly sliding forward or falling to one side.
Distribution
Nationwide domestic distribution. Foreign distribution to Canada, Costa Rica, Indonesia, South Korea, and Taiwan.
Initiated
Voluntary: Firm initiated
Recall number
Z-0824-2019

Terminated Class II Medical devices

Recall started
20 December 2018
Terminated
22 July 2019
Reason
The software in the systems may not process the cycle originally intended. This could result in damage to the various medical devices and instruments processed in the washer/disinfector or potential for inadequate cleaning or disinfection.
Distribution
The products were distributed to the following US states: AL, AR, CA, CO, FL, GA, IL, LA, MA, MD, ME, MI, NC, NJ, NM, NY, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, and WV. The products were distributed to the following foreign countries: Bahrain, Brazil, Canada, China, Saudi Arabia, South Korea,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0996-2019, Z-0995-2019

Terminated Class II Medical devices

Recall started
26 July 2018
Terminated
30 March 2020
Reason
There is the potential for a scenario where a punctured cup of sterilant expires while the system is powered down, upon power-up the unit will not detect the expired cup of sterilant; allowing for subsequent cycles to be run.
Distribution
Worldwide Distribution: US(nationwide) and to countries of: Albania, Algeria, Australia, Austria, Bahrain, Bangladesh, Belgium, Belarus, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany,
Initiated
Voluntary: Firm initiated
Recall number
Z-2959-2018

Terminated Class II Medical devices

Recall started
23 May 2018
Terminated
5 November 2019
Reason
The firm has become aware that the Reliance 1227 Cart and Utensil Washer/Disinfector s Chemical Low Level alarm, intended to prevent the user from initiating a cycle when a low chemical level situation occurs, does not operate as intended. Currently, if a low chemical level situation occurs, the alarm will only be generated at the unit s next power-up or when accessing service mode. The failure of the Reliance 1227 to identify that the chemicals used for cleaning are low or empty could result in bedpans and urinals, basins, case carts, beds, theater shoes and other miscellaneous reusable patient care items not being properly cleaned before disinfection or reuse.
Distribution
Worldwide Distribution - US Nationwide in the states of AL, AK, CA, CO, CN, DE, FL, GA, IL, IA, KS, KY, LA, MA, MI, NJ, NY, ND, OH, OK, OR, PA, RI, SC, TN, TX, VT, VA, WI, and WV. Foreign distribution to Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-2392-2018

Terminated Class II Medical devices

Recall started
28 September 2017
Terminated
12 September 2018
Reason
Correction to update sterilizer software to prohibit the use of expired sterilant prior to the start of a sterilization cycle.
Distribution
Distributed in 44 states: AL, AK, AZ, AR, CA, CO, DE, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, WY; Puerto Rico, and the following countries: Algeria, Australia, Azerbaijan, Bahrain, Brazil, Brunei
Initiated
Voluntary: Firm initiated
Recall number
Z-0772-2018

Terminated Class II Medical devices

Recall started
15 June 2017
Terminated
17 July 2018
Reason
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Distribution
US Nationwide Distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0749-2018, Z-0750-2018, Z-0754-2018, Z-0751-2018, Z-0743-2018, Z-0759-2018, Z-0770-2018, Z-0761-2018, Z-0767-2018, Z-0766-2018, Z-0763-2018, Z-0755-2018 and 17 more

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.