Recall Record

Sylmar, CA, United States

St Jude Medical Inc. recalls

12 recalls recorded since 2018, most recently on 3 February 2025. 7 are listed as ongoing. Status and classification are as published by the FDA.

Recalls12
Ongoing (FDA)7
Terminated or completed5
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
3 February 2025
Reason
Due to issues associated with data migration of patient information and results to a Cloud service, the firm has received complaint associated with some patient profile information being duplicated while other patient information is missing.
Distribution
U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA WI, and WV.
Initiated
Voluntary: Firm initiated
Recall number
Z-1427-2025

Ongoing Class II Medical devices

Recall started
6 November 2024
Reason
Due to a data migration from on-premises server to the cloud environment, some customers reported that they were not able to transmit readings.
Distribution
U.S.: CA, CT, GA, IL, MI, MN, NC, NJ, NY, OR, PA, TX, and WA. O.U.S.: N/A
Initiated
Voluntary: Firm initiated
Recall number
Z-0987-2025

Ongoing Class II Medical devices

Recall started
5 October 2024
Reason
As a result of a cloud data migration of Patient and Hospital Electronic Systems (PES and HES) reverted to an outdated configuration causing incorrect/inaccurate pulmonary artery (PA) pressure readings.
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NE, NH, NJ, NV, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI and the countries of Australia, and Canada.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0989-2025, Z-0988-2025

Ongoing Class II Medical devices

Recall started
30 April 2024
Reason
One lot of product has dilators that are too short and will not extend outside the introducer sheath.
Distribution
US and Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-2348-2024

Ongoing Class II Medical devices

Recall started
4 October 2023
Reason
Potential for damaged and frayed power connector plug with repeated bending or manipulation beyond 90 degrees.
Distribution
Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, IRELAND, ITALY, KUWAIT, NETHERLANDS, PORTUGAL, SAUDI ARABIA, SPAIN, SWEDEN, SWITZERLAND, UNITED ARAB EMIRATES, UNITED KINGDOM.
Initiated
Voluntary: Firm initiated
Recall number
Z-0218-2024

Ongoing Class II Medical devices

Recall started
7 February 2023
Reason
Select CardioMEMS PA Sensors (Model CM2000) operate outside of the intended frequency range at higher elevations when readings are taken over approximately 2,000 ft. (610 meters) above sea level; operation outside of the intended radiofrequency has the potential to result in inaccurate readings or sensor signal acquisition difficulties.
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1225-2023, Z-1223-2023, Z-1224-2023

Ongoing Class II Medical devices

Recall started
7 February 2023
Reason
Radiated Radiofrequency (RF) emissions of the CardioMEMS Patient Electronics Systems (PES) Models CM1000, CM1010 and CM1100 and CardioMEMS Hospital Electronics Systems (HES) Model CM3000 exceed certain emissions limits and have the potential to cause interference with other medical devices, such as implantable neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors when in close proximity to an active Hospital or Patient Electronics System (i.e., during readings).
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Canada, Colombia, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Kuwait, Netherlands, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, UK, United Arab Emirates.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1221-2023, Z-1219-2023, Z-1218-2023, Z-1220-2023

Terminated Class I Medical devices

Recall started
20 June 2019
Terminated
3 April 2024
Reason
Electrical failures were identified in cardioverter defibrillators (ICDs) due to damaged aluminum wires. ICDs may contain electrical wire connections which may not be completely insulated. The potential patient impact could be the inability to deliver high voltage therapy. There is no available option to verify the vulnerability status for implanted devices.
Distribution
Worldwide distribution. US nationwide, and countries: DE, FR, IT, ES, NL, PT, SE, HU, DK, PL, IL, GP, DZ, CH, SA, SK, LU, NZ, VN, and NO
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2071-2019, Z-2073-2019, Z-2074-2019, Z-2072-2019, Z-2075-2019, Z-2070-2019

Terminated Class II Medical devices

Recall started
11 October 2018
Terminated
13 September 2022
Reason
The most proximal unsegmented electrode of the Deep Brain Stimulation leads, may be constructed with MP35N instead of the required 90/10 platinum-iridium.
Distribution
Worldwide distribution - US nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, MD, MN, MO, MT, NC, NJ, NY, OH, PA, and WA. International distribution in the countries of Argentina, Australia, Austria, Belgium, Brazil, Colombia, Czech Republic, Finland, France, Germany, Hungary, Italy,
Initiated
Voluntary: Firm initiated
Recall number
Z-0593-2019

Terminated Class II Medical devices

Recall started
16 April 2018
Terminated
8 March 2024
Reason
The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
Distribution
U.S. Nationwide distribution
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2329-2018, Z-2326-2018, Z-2320-2018, Z-2321-2018, Z-2328-2018, Z-2325-2018, Z-2322-2018, Z-2319-2018, Z-2318-2018, Z-2323-2018, Z-2327-2018, Z-2324-2018

Terminated Class II Medical devices

Recall started
16 April 2018
Terminated
4 February 2021
Reason
The firm is advising physicians that exposure to sub-freezing temperatures during the supply chain process caused a transient battery voltage drop for a small number of Confirm Rx Model DM3500 Insertable Cardiac Monitoring (ICM) devices. This drop is an expected and normal behavior for this battery chemistry, but causes an incorrect display of a low battery indicator even after the battery voltage returns to normal.
Distribution
U.S. Nationwide Distribution: PR, MI, NJ, IL, FL, AR, PA, KY, CA, TN, IN, AL, TX, AZ, SC, OK, VA, NH, NM, WA, CO, IA, WI, NV, UT, OH, NC, MO, DE, WV, MD, OR, NE, VT, KS, MA, IA, WY, CT, GA, MS, LA, DC, LA,
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2334-2018, Z-2335-2018

Terminated Class II Medical devices

Recall started
9 March 2018
Terminated
3 November 2020
Reason
The firm received complaints of error messages that occurred during routine impedance checks on Proclaim DRG IPGs, model 3664. Some complaints were also associated with transient over stimulation which created discomfort for the patients.
Distribution
US
Initiated
Voluntary: Firm initiated
Recall number
Z-1170-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.