Recall Record

Sylmar, CA, United States

St. Jude Medical, Cardian Rhythm Management Division recalls

1 recall recorded since 2018, most recently on 15 March 2021. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)1

Ongoing Class I Medical devices

Recall started
15 March 2021
Reason
Due to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the device header and feedthrough
Distribution
Worldwide distribution: U.S (nationwide) to states including.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.( countries): [8
Initiated
Voluntary: Firm initiated
Recall numbers
Z-1531-2021, Z-1530-2021