Spacelabs Healthcare, Inc. recalls

15 recalls recorded since 2018, most recently on 4 June 2026. 10 are listed as ongoing. Status and classification are as published by the FDA.

Recalls15
Ongoing (FDA)10
Terminated or completed5
Class I (FDA)1

Ongoing Class II Medical devices

Recall started
4 June 2026
Reason
Due to off-the-shelf display port component that has audio/video failures
Distribution
Worldwide - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of CANADA, FRANCE, ITALY, MALAYSIA, THAILAND.
Initiated
Voluntary: Firm initiated
Recall number
Z-2916-2026

Ongoing Class II Medical devices

Recall started
26 March 2026
Reason
Due to a manufacturing issue (i.e. malfunctioning of electrical Leakage tester).
Distribution
Worldwide - US Nationwide di GA, KS, MI, OH, SC, TX, WA and the country of ARGENTINA.
Initiated
Voluntary: Firm initiated
Recall number
Z-2052-2026

Ongoing Class II Medical devices

Recall started
27 February 2026
Reason
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
Distribution
Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1793-2026

Ongoing Class II Medical devices

Recall started
16 September 2025
Reason
Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
Distribution
Worldwide - US Nationwide distribution in the state of MN and the countries of CA, DE, FR, GB, IT, NC.
Initiated
Voluntary: Firm initiated
Recall number
Z-0416-2026

Ongoing Class II Medical devices

Recall started
5 September 2025
Reason
Due to display cables used for monitors were not in compliance with electromagnetic interference from inadequately shielded cables which may pose serious risk to patient's safety by interfering with the operation of another electronically sensitive medical device are in use.
Distribution
Worldwide - United States Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WY and the countries of AFG, ARE, ARG, AUS, AUT, BG
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0169-2026, Z-0168-2026, Z-0167-2026

Ongoing Class II Medical devices

Recall started
25 November 2024
Reason
Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
Distribution
U.S.: AR, CA, CO, DE, FL, FA, LA, MA, MI, MN, MT, NC, NE, NU OH, OR, PA, Puerto Rico, TN, TX, UT, VA, WA, and WI. O.U.S.: Argentina Denmark Kuwait Serbia Australia Dominican Republic Lebanon Singapore Austria Ecuador Lithuania Spain Belgium El Salvador Martinique Sweden Benin France Mexico Switzerl
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0684-2025, Z-0685-2025

Ongoing Class II Medical devices

Recall started
25 November 2024
Reason
Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
Distribution
US: OK, MN, IA, SC, NE, MI, NY, KY, GA, MO, MT, LA, WA, SD, CO, PA, WY, TX, ID, NJ, FL, OH, MS, OR, AK, IL, AR, KS, AL, TN, WI, VA, NM, CA, IN, RI, NV, MA, NC, ME, CT, AZ, MD, PR. OUS: Kuwait, Australia, Canada, Roumanie, Greece, Philippines, Spain, Italy, Hong Kong, Switzerland, Panama, Taiwan, Ne
Initiated
Voluntary: Firm initiated
Recall number
Z-0809-2025

Ongoing Class II Medical devices

Recall started
25 November 2024
Reason
Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.
Distribution
US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL, WA, MA, WI, OR, MS, ID, NY, KS, PR. OUS: Canada, Taiwan, Czech Republic, France, Greece, Italy, Kuwait, Philippines, Poland.
Initiated
Voluntary: Firm initiated
Recall number
Z-0810-2025

Ongoing Class II Medical devices

Recall started
12 April 2023
Reason
Command modules will display the following when parameter processing is suspended, depending on software version: 1) Older versions will have waveforms present with no alarms, 2) Newer versions have waveforms absent with no alarms. This may cause clinicians to miss a potentially catastrophic event.
Distribution
US Nationwide distribution in the states of SD, KY, MO, WA, OH, NC, CA, OR, TX, GA, IL, OK, SC, PA, AL, KS, FL, ID, MI, MN, IA, NJ, MD, VA, AR, CT, NE, LA, AK, MS, ME, WV, NY, AZ, TN, WY, IN, MA, MT, CO, DE, ND, WI, NV, RI, NH, UT.
Initiated
Voluntary: Firm initiated
Recall number
Z-1503-2023

Ongoing Class II Medical devices

Recall started
31 May 2022
Reason
Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.
Distribution
Worldwide distribution - US Nationwide distribution in the states of NY, AR, CO, FL, NJ, AL, KY, TN, SC, WA, OK, MN, PA, GA, VA, CA, OH, IA, KS, MI, IN, TX, MT, MO, ME, LA, IL, SD, MA, OR, ID, MD, AZ, AK and the countries of MB, ON, FR, QC, BC, IS, CT, RM.
Initiated
Voluntary: Firm initiated
Recall number
Z-1513-2022

Terminated Class II Medical devices

Recall started
6 November 2020
Terminated
18 October 2024
Reason
During system upgrade incorrect comments may be added to existing tests. In such instance: Resting ECG and ABP tests which did not contain comments at the time of the upgrade, may have unrelated comments added. Existing comments will not be overwritten or removed. Rhythm ECG Tests may have comments overwritten with unrelated comments. Any tests created after the upgrade are not affected.
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of PR, MN, NE, FL, IL, CA, NY, ND, AR, WI, SD, TN, MO, MA, TX, MT, OH, MI, IA, NJ, CT, NC, UT, WV, GA, CO, PA, AL, DC, NV, DE, WA, OR, IN, SC, HI, MS, KS, ID, MD, LA, NH, VA, ME, AK, VT, OK, AZ, KY, NM, RI and the countrie
Initiated
Voluntary: Firm initiated
Recall number
Z-0563-2021

Terminated Class II Medical devices

Recall started
30 August 2019
Terminated
18 April 2024
Reason
The firm received reports of loss of audio alarm after a power failure or cable disconnection. The loss of audio for a telemetry patient could delay the recognition of an alarm condition.
Distribution
Worldwide distribution - US Nationwide distribution including in the 40 states and Puerto Rico, and multiple countries including Canada.
Initiated
Voluntary: Firm initiated
Recall number
Z-2261-2020

Terminated Class II Medical devices

Recall started
28 January 2019
Terminated
24 February 2022
Reason
Several reports were received that patient records were printed with the correct patient demographics but containing another patient's waveforms.
Distribution
Distribution was nationwide. There was also government and military distribution. Foreign distribution was made to Argentina, Bahrain, Bolivia, Brazil, Canada, Cayman Islands, Chile, China, Costa Rica, Czech Republic, France, Hong Kong, India, Indonesia, Italy, Kuwait, Malaysia, Mexico, Netherla
Initiated
Voluntary: Firm initiated
Recall number
Z-2035-2019

Terminated Class I Medical devices

Recall started
11 July 2018
Terminated
20 March 2024
Reason
Arkon Anesthesia Delivery System may go into a failed state (mechanical ventilation ceases) while the machine is in use or while idle.
Distribution
Worldwide Distribution: U.S (nationwide) to states of.: LA, NE, CO, WY, MT, NM, IN, ME, NY, IL, MA, TX, PA, PR; and countries (OUS) of: PERU BOLIVIA, PANAMA, MALAYSIA, POLAND, IRELAND, MOROCCO, MEXICO, CANADA, UNITED KINGDOM, FRANCE, and CHINA, ECUADOR.
Initiated
Voluntary: Firm initiated
Recall number
Z-2589-2018

Terminated Class II Medical devices

Recall started
17 January 2018
Terminated
29 November 2018
Reason
Spacelabs has received multiple complaints reporting telemetry beds dropping off the Spacelabs Central Station resulting in a loss of monitoring.
Distribution
US Distribution to the states of : VA, NJ, MI, GA, and Internationally to France
Initiated
Voluntary: Firm initiated
Recall number
Z-0532-2018

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.