Southmedic, Inc. recalls

4 recalls recorded since 2018, most recently on 27 October 2022. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)2
Terminated or completed2
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
27 October 2022
Reason
Upon review of its batch record, Southmedic identified that lot # W92761 was not sent out for sterilization, and was instead transferred directly to distribution where it was sold and distributed to customers as non-sterile product labelled as sterile.
Distribution
US Nationwide distribution in the state of IL.
Initiated
Voluntary: Firm initiated
Recall number
Z-0279-2023

Ongoing Class II Medical devices

Recall started
25 May 2021
Reason
Potential for oxygen tubing to disconnect from mask or loosen. Disconnected tubing results in a loss of device function and use of the device could result in inappropriate or a lack of oxygen delivery to the patient.
Distribution
U.S.: IL O.U.S.: N/A
Initiated
Voluntary: Firm initiated
Recall number
Z-2141-2021

Terminated Class II Medical devices

Recall started
6 November 2017
Terminated
2 April 2020
Reason
Scalpels lot contains non-sterile products labeled as sterile.
Distribution
US Distribution to the states of : IA, IL, NE, NY and TX
Initiated
Voluntary: Firm initiated
Recall number
Z-1892-2018

Terminated Class II Medical devices

Recall started
17 October 2017
Terminated
29 March 2024
Reason
Potential for loose or missing CO2 diffusers which could result in a non-functioning mask.
Distribution
US Distribution to the state of California, and Internationally to Canada and Sweden
Initiated
Voluntary: Firm initiated
Recall number
Z-1288-2018