SIGHT DIAGNOSTICS LTD recalls
4 recalls recorded since 2018, most recently on 13 May 2024. 3 are listed as ongoing. Status and classification are as published by the FDA.
Recalls4
Ongoing (FDA)3
Terminated or completed1
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 13 May 2024
- Reason
- One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, FL, MS, TX, WI and the countries of Canada, Israel, Slovenia, Trinidad & Tobago.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2173-2024
- Sight OLO CBC Test Kit, Model/Catalog Number: TK1. A disposable cartridge and sample preparation tools for use with the Sight OLO CBC analyzer.
Ongoing Class II Medical devices
- Recall started
- 14 June 2022
- Reason
- The default reference ranges for MCH, MCHC, and RDW for age group "Child 2-11y" are mixed up and incorrectly displayed on OLO devices running software version 2.63, 2.63.1, and 2.63.1.1. Due to the issue, results may incorrectly appear to be within or outside the reference range.
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of Alabama, Arizona, California, Florida, Idaho, Illinois, Kansas, Kentucky, Massachusetts, Michigan, Minnesota, Mississippi, New Jersey, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Washington, Wisconsin and the c
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1469-2022
- Sight OLO, Automated Hematology Analyzer, Catalog No. OLO-U1
Ongoing Class II Medical devices
- Recall started
- 20 January 2022
- Reason
- In instances where custom reference ranges were configured on the device post installation, a possibility to inadvertently apply changes to reference range values was found on software versions 2.59.3 and all earlier versions, which can potentially lead to displaying and printing incorrectly-configured reference range values
- Distribution
- Worldwide distribution - US Nationwide and the countries of AUSTRALIA, CANADA, CHILE, CURACAO, FINLAND, GEORGIA, GERMANY, GREECE, HONG KONG, IRELAND, ISRAEL, ITALY, MEXICO, NETHERLANDS, PORTUGAL, SINGAPORE, SLOVENIA, UNITED ARAB EMIRATES, UNITED KINGDOM.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-0963-2022
- The Sight OLO device is a computer vision based platform for blood analysis. The platform combines computer-vision algorithms for image processing to identify and quantify blood components (e.g., red blood cells) and t
Terminated Class II Medical devices
- Recall started
- 17 March 2021
- Terminated
- 8 June 2022
- Reason
- Discrepancy in labeling between the adult CBC reference ranges presented in the OLO Operator's Manual and the OLO software
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of Alabama, Florida, Massachusetts, Minnesota, Missouri, New York, Ohio, Texas, Virginia, Washington and the countries of AUSTRALIA, CHILE, UNITED KINGDOM, GEORGIA, ISRAEL, Mexico, NETHERLANDS, Portugal, Russia, Singapore, Thailand.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-2517-2021
- Sight OLO-E1/U1 Software and Operators Manual Software v2.56.1 Catalog #: 1 OLO-U1 Catalog #: 2 OLO-E1 revision 1.9, revision 1.10
A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.