SIGHT DIAGNOSTICS LTD recalls

4 recalls recorded since 2018, most recently on 13 May 2024. 3 are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)3
Terminated or completed1
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
13 May 2024
Reason
One Test Kit lot (TK1-10114) shows a positive bias in the platelet count due to bacterial contamination which can result in elevated counts with a bias of 7.5 k/ul. This results in the test kit lot performing outside of the device specification.
Distribution
Worldwide - US Nationwide distribution in the states of AL, FL, MS, TX, WI and the countries of Canada, Israel, Slovenia, Trinidad & Tobago.
Initiated
Voluntary: Firm initiated
Recall number
Z-2173-2024

Ongoing Class II Medical devices

Recall started
14 June 2022
Reason
The default reference ranges for MCH, MCHC, and RDW for age group "Child 2-11y" are mixed up and incorrectly displayed on OLO devices running software version 2.63, 2.63.1, and 2.63.1.1. Due to the issue, results may incorrectly appear to be within or outside the reference range.
Distribution
Worldwide distribution - US Nationwide distribution in the states of Alabama, Arizona, California, Florida, Idaho, Illinois, Kansas, Kentucky, Massachusetts, Michigan, Minnesota, Mississippi, New Jersey, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Washington, Wisconsin and the c
Initiated
Voluntary: Firm initiated
Recall number
Z-1469-2022

Ongoing Class II Medical devices

Recall started
20 January 2022
Reason
In instances where custom reference ranges were configured on the device post installation, a possibility to inadvertently apply changes to reference range values was found on software versions 2.59.3 and all earlier versions, which can potentially lead to displaying and printing incorrectly-configured reference range values
Distribution
Worldwide distribution - US Nationwide and the countries of AUSTRALIA, CANADA, CHILE, CURACAO, FINLAND, GEORGIA, GERMANY, GREECE, HONG KONG, IRELAND, ISRAEL, ITALY, MEXICO, NETHERLANDS, PORTUGAL, SINGAPORE, SLOVENIA, UNITED ARAB EMIRATES, UNITED KINGDOM.
Initiated
Voluntary: Firm initiated
Recall number
Z-0963-2022

Terminated Class II Medical devices

Recall started
17 March 2021
Terminated
8 June 2022
Reason
Discrepancy in labeling between the adult CBC reference ranges presented in the OLO Operator's Manual and the OLO software
Distribution
Worldwide distribution - US Nationwide distribution in the states of Alabama, Florida, Massachusetts, Minnesota, Missouri, New York, Ohio, Texas, Virginia, Washington and the countries of AUSTRALIA, CHILE, UNITED KINGDOM, GEORGIA, ISRAEL, Mexico, NETHERLANDS, Portugal, Russia, Singapore, Thailand.
Initiated
Voluntary: Firm initiated
Recall number
Z-2517-2021

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.