Siemens AG/Siemens Healthcare GmbH recalls
1 recall recorded since 2018, most recently on 26 June 2024. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 26 June 2024
- Reason
- A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.
- Distribution
- Worldwide - US Nationwide distribution in the states of TX, MA, NH, MN, OK, NJ, SC, NY, MI, IL, MO, NC.
- Initiated
- FDA Mandated
- Recall numbers
- Z-2241-2024, Z-2242-2024, Z-2240-2024
- ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.