Sentinel CH SpA recalls

4 recalls recorded since 2018, most recently on 29 September 2023. 2 are listed as ongoing. Status and classification are as published by the FDA.

Recalls4
Ongoing (FDA)2
Terminated or completed2
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
29 September 2023
Reason
A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.
Distribution
Worldwide - US Nationwide distribution.
Initiated
Voluntary: Firm initiated
Recall numbers
Z-0325-2024, Z-0326-2024

Terminated Class III Medical devices

Recall started
29 July 2021
Terminated
24 July 2023
Reason
Calibrator vials were incorrectly labeled. Use of incorrectly labeled vials will cause the instrument to generate a warning that calibrators are incorrectly loaded and calibration will not be completed successfully.
Distribution
US Nationwide distribution in the states of LA, MD, MN, ND, NY, WV, and Puerto Rico.
Initiated
Voluntary: Firm initiated
Recall number
Z-0033-2022

Ongoing Class III Medical devices

Recall started
30 April 2021
Reason
Claim for stability after reconstitution from lyophilized to liquid form does not meet the requirements. New IFU instructions for handling the product provided to end users.
Distribution
Worldwide distribution - US Nationwide and the countries of AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, CANADA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE, IRELAND, ISRAEL, ITALY, JORDAN, NEW ZEALAND, POLAND, PORTUGAL, SERBIA, SOUTH AFRICA, SPAIN, SWEDEN, THE NETHERLANDS,
Initiated
Voluntary: Firm initiated
Recall number
Z-2012-2021

Terminated Class II Medical devices

Recall started
28 June 2019
Terminated
24 July 2020
Reason
Added an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential for carryover from the ARCHITECT Lactate Dehydrogenase (LDH) reagent (LN 2P56) may result infalsely elevated lithium patient results
Distribution
Nationwide Foreign: ARGENTINA AUSTRALIA AUSTRIA AZERBAIJAN BANGLADESH BELARUS BELGIUM BOSNIA AND HERZEGOVI BRAZIL BRUNEI CANADA CHILE COLOMBIA COSTA RICA CROATIA CZECH REPUBLIC DENMARK FAROE ISLANDS FINLAND FRANCE GERMANY GREECE GREENLAND HONG KONG IRELAND ISRAEL ITALY J
Initiated
Voluntary: Firm initiated
Recall numbers
Z-2551-2019, Z-2552-2019

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.