Recall Record

Hamburg, N/A, Germany

Seca recalls

1 recall recorded since 2018, most recently on 19 June 2019. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
19 June 2019
Reason
It has been determined that the wheels on the baby scale cart can unexpectedly loosen and in the worst case, detach from the cart completely. If this occurs the cart can become unstable and tip over, potentially leading to the scale falling from the cart. If a baby is left unattended in a cart, the baby could fall off the scale and be subjected to several serious injuries.
Distribution
US: CA, MI, WA, WV, VA, NM, LA, TN, IL, PA, OH, FL, WA, OR, NY, CO, IA, MN, TX, MA, NC, LA, KY, NM AL,DC, NJ, AK, AZ, NV, SC, HI, MD, WI, RI, DE, MO OUS: Canada
Initiated
Voluntary: Firm initiated
Recall number
Z-1429-2020

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.