Recall Record

Denzlingen, N/A, Germany

Scholly Fiberoptic Gmbh recalls

1 recall recorded since 2018, most recently on 23 June 2023. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
23 June 2023
Reason
There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.
Distribution
Nationwide Foreign: Japan, South Korea
Initiated
Voluntary: Firm initiated
Recall number
Z-2364-2023

A terminated FDA recall means the FDA has determined that all reasonable efforts to remove or correct the product have been made. What statuses mean.