Houston, TX, United States

Saranas, Inc. recalls

1 recall recorded since 2018, most recently on 10 August 2022. 1 is listed as ongoing. Status and classification are as published by the FDA.

Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0

Ongoing Class II Medical devices

Recall started
10 August 2022
Reason
Bleed monitoring system device design error, causing the device to not power on and complete the series of self-tests so instructions for use being updated to include device check before use ensure that the device is functioning appropriately.
Distribution
US Nationwide distribution including in the states of PA, MI, FL, WV, TX, NJ, GA, LA, NY, MA, CT, AZ, OH, TN, VA.
Initiated
Voluntary: Firm initiated
Recall number
Z-1684-2022