Houston, TX, United States
Saranas, Inc. recalls
1 recall recorded since 2018, most recently on 10 August 2022. 1 is listed as ongoing. Status and classification are as published by the FDA.
Recalls1
Ongoing (FDA)1
Terminated or completed0
Class I (FDA)0
Ongoing Class II Medical devices
- Recall started
- 10 August 2022
- Reason
- Bleed monitoring system device design error, causing the device to not power on and complete the series of self-tests so instructions for use being updated to include device check before use ensure that the device is functioning appropriately.
- Distribution
- US Nationwide distribution including in the states of PA, MI, FL, WV, TX, NJ, GA, LA, NY, MA, CT, AZ, OH, TN, VA.
- Initiated
- Voluntary: Firm initiated
- Recall number
- Z-1684-2022
- Early Bird Bleed Monitoring System Introducer Sheath Set, REF: FG001-01 (6F), FG001-02 (8F)